7. 

Initial Conditions and Data Entries

•  Logged readings include initial conditions and data entries. 

The initial conditions are the thermocouple type and the

offsets for each thermocouple input. You can only view initial

conditions using the included software. 

•  The data entries are a time stamp and the T1, T2 or T1-T2 reading. 

You can view these values by pressing “RECALL (SETUP)” 

momentarily or using the included software. The thermometer 

has 9,999 memory locations. The thermometer stores 9,999 sets 

of temperature readings and one set of initial conditions when 

logging continuously. It stores 9,999 sets of temperature readings 

and initial conditions when logging individual points manually. 

8. 

Starting and Stopping Logging

•  Memory clear, and PC communications are inaccessible during

logging. Recall function is enabled during logging.

•  Set the time and the logging interval, see: Set Up Instructions

sections, E: Time, and H: Change Logging Interval.

Change Logging Interval

Change Logging Interval

•  Press “LOGGING (CLEAR)” momentarily to start logging. The 

display shows “ LOG “.

•  Press “LOGGING (CLEAR)” momentarily again to stop logging.

9. 

Clearing Memory

•  When memory is full, “ 

FULL

 “ appears in the secondary

display and logging stops. You can clear memory in normal or

MAX/MIN mode.

•  Press “LOGGING (CLEAR)” for more than 1 second to display 

“ 

 ” in the RECALL mode.

•  Press “LOGGING (CLEAR)” for more than 1 second again to 

delete all logged readings from memory.

•  Press “EXIT” momentarily or turn off the thermometer to stop 

clearing memory.

10.  Viewing Logged Readings

•  Press “RECALL” momentarily to display the last logged

reading. If there is no logged readings, “   ” appears in

the secondary display, and the thermometer returns to the 

previous mode.

•  Press and Hold “  ” or “

” to scroll through the logged

readings. The display shows each logged reading and its time

stamp, which is displayed 2 seconds after its memory location

appeared in the secondary display.

•  Press “MAX/MIN” momentarily to step through the minimum, 

maximum, average, and currently logged reading.

•  Press “T1/T2” momentarily to display the logged readings you

want to view. Press “T1-T2” for more than 1 second to display 

differential values between the logged readings you want to 

view.

•  Press “RECALL” momentarily or turn off the thermometer to 

stop viewing logged readings.

•  NOTE: The thermometer calculates the minimum and 

NOTE:

NOTE:

maximum of all logging sessions in memory.

11.  USB Port / External Power

•  This thermometer is equipped with an USB interface port and

cable. Optional Windows® compatible software is available

for download at www.kleintools.com.  This kit is required to

connect the thermometer to a PC. 

•  The kit allows you to transfer the current data measured and 

the contents of the thermometer’s memory to a PC. Refer to

the “HELP” menu in the software for further details.

•  The DTL302 can be powered through the USB cable or

included AC adapter. This allows long term monitoring or

logging without consuming battery power

•  NOTE: The AC adapter and/or USB cable does NOT recharge 

NOTE:

NOTE:

the installed batteries.

Содержание DTL302

Страница 1: ...PERATURE INPUTS IP67 WATER DUSTPROOF RATING MULTI PARAMATER BACKLIT DISPLAY CONVENIENT PROBE STORAGE MIN MAX DATA HOLD EXTERNAL USB POWER FOR LONG TERM LOGGING USB LOGGING SOFTWARE DIFFERENTIAL T1 T2 MAGNETIC MOUNT STRAP ENGLISH Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...

Страница 2: ...2 ...

Страница 3: ...tion CATEGORY I Pollution Degree 2 IEC 1010 1 WARNINGS WARNINGS To ensure safe operation and service of the tester follow these instructions Failure to observe these warnings can result in severe injury or death Follow the manufacturer s maintenance procedures when servicing equipment Place ONLY thermocouples type K J T or E in the thermocouple ports Make sure your meter is set for the proper ther...

Страница 4: ...ress to turn the backlight on and off The backlight turns off after 30 seconds without any button pressed If the battery is low the backlight is disabled D Max Min Press to step through the maximum minimum and average readings in the secondary display When viewing logged readings the maximum minimum and average measurements are displayed Press and Hold for more than 1 second to turn off this displ...

Страница 5: ...clear logged data press and hold until appears in the lower display To execute the clear press and hold until appears The ENTER function is activated in the SETUP mode only DISPLAY 1 2 3 4 5 6 7 12 8 9 10 11 1 Primary Display T1 T2 reading 2 TYPE KJET Type of thermocouple 3 SETUP Setup is in progress 4 RECALL The logged readings are displayed 5 OFFSET The thermocouple measurement includes an offse...

Страница 6: ...me must be set during the same session NOTE NOTE Both the date and time must be set in order to use the NOTE Both the date and time must be set in order to use the NOTE logging feature D Time Unit While in SETUP mode scroll until the display shows in the primary display and in the secondary display Press ENTER to indicate you want to set the time units The display shows h m blinking Press or to se...

Страница 7: ...press ENTER to select Holding down or causes the number to change more quickly I Setting Temperature Offset You can adjust the thermometer s readings to compensate for the errors of a specific thermocouple The allowable compensation range is 5 0 C or K and 9 0 F The DTL302 allows you to store individual offsets for T1 and T2 Press or until the display shows OFFSET Press ENTER to display T1 reading...

Страница 8: ...inimum MAX maximum or the AVG average readings in the secondary display The elapsed time since entering MAX MIN mode appears with the average reading The time at which the MINIMUM or MAXIMUM occurred appears with the MINIMUM or MAXIMUM reading on the display Press T1 T2 T1 T2 momentarily to toggle showing the maximum MAX of T1 or T2 readings and its displayed time Press T1 T2 T1 T2 for more than 1...

Страница 9: ... off the thermometer to stop clearing memory 10 Viewing Logged Readings Press RECALL momentarily to display the last logged reading If there is no logged readings appears in the secondary display and the thermometer returns to the previous mode Press and Hold or to scroll through the logged readings The display shows each logged reading and its time stamp which is displayed 2 seconds after its mem...

Страница 10: ... add 0 2 of reading T Type 0 3 2 7 F 1 5 C 148 F 100 C add 0 5 of reading Temperature Coefficient 0 01 of reading 0 1 F per F 0 05 C per C for 64 F 18 C or 82 F 28 C Below 148 F 100 C add 0 05 of reading for J K and E type and 0 1 of reading for T type Real Time Clock Tolerance About 1 second per day Max Differential Common Mode Voltage 1V Maximum voltage difference between any pair of inputs Temp...

Страница 11: ...11 ...

Страница 12: ... Always dispose of old batteries promptly in a manner consistent with local disposal regulations Place a fresh batteries in the compartment NOTE If you do not plan to use the meter for a month or more remove the battery and store it in an area that won t be damaged by a leaking battery Reattach the battery compartment cover to the meter and reinstall the screws CLEANING Turn instrument off and dis...

Отзывы: