Scan Recipes - Creating and Editing a Scan Recipe
KLA-Tencor P-16+ / P-6 User’s Guide
3-24
KLA-Tencor Confidential
0142530-000 AB
3/13/09
Traces
This assigns the number of scans that are made in the X-direction across the
Y Scan
Size
direction. In
, the number in the
Traces
variable box would be
8
.
If the
Y Scan Size
is set [Y Scan Size = (Traces -1) x Y Spacing], when the
Traces:
parameter is entered, the
Y Spacing
parameter automatically adjusts to reflect the
appropriate spacing between scans.
Setting the Number of
Traces:
To change the number of
Traces
in a 3D scan,
highlight the current
Traces
value and type in the new number of traces. (See also
Automatic Parameter Adjustment
Y
Spacing (
μ
m)
This variable sets the distance in the Y-direction between X-direction scan traces in a
3D scan. The spacing is very important to final 3D data collection set because,
together with the stylus radius, it determines the essential resolution of the feature that
is scanned.
Consider to following example: If the distance between scans is too great with respect
to the stylus radius, important variations in the scanned feature might be missed.
Figure 3.19
Traces - Scan Perimeter with Traces
X Scan Size
Y Scan Size
This scan contains 8 traces,
illustrated as dotted line arrows.
The number of traces is defined
in the
Traces
parameter.
Feature being scanned
NOTE:
The first trace occurs at y = 0 µm. Setting the number of
traces to an odd number ensures that the y-spacing will be an even
number. For example, a 100 µm y-size and 11 traces produces 10 µm
between each trace.
Содержание P-16+
Страница 4: ...KLA Tencor P 16 P 6 User s Guide KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 24: ...KLA Tencor P 16 P 6 User s Guide LOT 4 KLA Tencor Confidential 0142530 000 AB Friday March 13 2009 Preliminary...
Страница 30: ...Introduction Safety Safety KLA Tencor P 16 P 6 User s Guide 1 6 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 180: ...XY View Screen Aligning the Sample KLA Tencor P 16 P 6 User s Guide 4 32 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 196: ...View Scan Window Aborting A Scan KLA Tencor P 16 P 6 User s Guide 5 16 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 434: ...Backup and Restore KLA Tencor P 16 P 6 User s Guide 16 8 KLA Tencor Confidential 0241110 000 AB 3 13 09...
Страница 444: ...KLA TENCOR P 16 P 6 USER S GUIDE REVISION HISTORY 2 KLA TENCOR CONFIDENTIAL 0142530 000 AB 3 13 09...