Stylus Change Procedure - Stylus Removal and Replacement
3.
Place the
ProCal Wafer
precisely in the center of the stage, squarely positioned
with respect to the XY axis.
4.
Turn the vacuum ON using the switch on the upper left inside door frame.
5.
Close the stage door.
6.
Click OK in the message box. (See
.)
The
Scan Offset Calibration Option
dialog box is displayed (see
) on
top of the Calibration screen.
Two columns present the two options used to set up the Scan Offset Calibration.
The first column is the
Size
column. It is used to determine the length of the step
that is to be scanned and, therefore, which triangle the scan is to be performed
on. If the step is 150
μ
m, the system uses the 300
μ
m triangle. If the step is
500
μ
m, the system uses the 1000
μ
m (1 mm) triangle.
7.
Choose
150
μ
m
(standard) to continue with the calibration. (See
)
Figure 13.15
Manual Load from the Scan Offset Calibration Window
After loading the
ProCal Wafer, click
MAN
LOAD
again to send the
stage back under the stylus.
NOTE:
The Vacuum menu in the screen’s menu bar is not functional.
It does not effect the stage vacuum.
Figure 13.16
Scan Position Offset Calibration Options dialog box
Custom
or
Default
Click on
150
μ
m
.
Содержание P-16+
Страница 4: ...KLA Tencor P 16 P 6 User s Guide KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 24: ...KLA Tencor P 16 P 6 User s Guide LOT 4 KLA Tencor Confidential 0142530 000 AB Friday March 13 2009 Preliminary...
Страница 30: ...Introduction Safety Safety KLA Tencor P 16 P 6 User s Guide 1 6 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 180: ...XY View Screen Aligning the Sample KLA Tencor P 16 P 6 User s Guide 4 32 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 196: ...View Scan Window Aborting A Scan KLA Tencor P 16 P 6 User s Guide 5 16 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 434: ...Backup and Restore KLA Tencor P 16 P 6 User s Guide 16 8 KLA Tencor Confidential 0241110 000 AB 3 13 09...
Страница 444: ...KLA TENCOR P 16 P 6 USER S GUIDE REVISION HISTORY 2 KLA TENCOR CONFIDENTIAL 0142530 000 AB 3 13 09...