![KLA Tencor P-16+ Скачать руководство пользователя страница 169](http://html1.mh-extra.com/html/kla-tencor/p-16/p-16_user-manual_1974647169.webp)
0142530-000 AB
KLA-Tencor Confidential
4-21
3/13/09
KLA-Tencor P-16+ / P-6 User’s Guide
XY View Screen - Using Die Grid Navigation (Optional Feature, P-16+ only)
10.
Click
OK
when all variables are correct.
11.
Teach the upper left inside corner of the die: (follow the instructions on the
screen)
It is recommended to enter the die height and width from stepper mask
layout information for optimal die grid accuracy.
Position the die image using the arrow buttons so that the upper left corner
of the die is in the field of view.
Position the mouse cursor at the left inside corner of the die, as indicated in
the
illustration, and click.
Figure 4.16
Teach Die Grid - Teach First Position
The graphic represents
a single die. The thick gray
border is the area between the
various dies on the wafer. Teach
the first position as indicated by
the arrow in the illustration.
Notice the message,
instructing the operator.
Содержание P-16+
Страница 4: ...KLA Tencor P 16 P 6 User s Guide KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 24: ...KLA Tencor P 16 P 6 User s Guide LOT 4 KLA Tencor Confidential 0142530 000 AB Friday March 13 2009 Preliminary...
Страница 30: ...Introduction Safety Safety KLA Tencor P 16 P 6 User s Guide 1 6 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 180: ...XY View Screen Aligning the Sample KLA Tencor P 16 P 6 User s Guide 4 32 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 196: ...View Scan Window Aborting A Scan KLA Tencor P 16 P 6 User s Guide 5 16 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 434: ...Backup and Restore KLA Tencor P 16 P 6 User s Guide 16 8 KLA Tencor Confidential 0241110 000 AB 3 13 09...
Страница 444: ...KLA TENCOR P 16 P 6 USER S GUIDE REVISION HISTORY 2 KLA TENCOR CONFIDENTIAL 0142530 000 AB 3 13 09...