![KLA Tencor P-16+ Скачать руководство пользователя страница 159](http://html1.mh-extra.com/html/kla-tencor/p-16/p-16_user-manual_1974647159.webp)
0142530-000 AB
KLA-Tencor Confidential
4-11
3/13/09
KLA-Tencor P-16+ / P-6 User’s Guide
XY View Screen - Focusing the View
3.
Ensure that the zoom position in the dialog box
Zoom
field agrees with the
Zoom
Position
in the screen display. (See
4.
Click the
Save Zoom Position
check box. (See
5.
Click
OK
to save the position and lock the zoom function. (See
F
OCUSING
THE
V
IEW
4
Introduction
4
After the null and focus procedure, if the sample surface is not in focus, the focus
knobs can be used to bring the surface into focus. (This should only be required after
stylus change.)
The purpose of focusing the view is to sharpen the image in the video window. If the
focus is clear the first time, and the sample is flat, focus should be maintained each
time the stage moves to another location on the same sample surface.
Focus the Optics – Top- or Side-View
4
1.
Raise the measurement head.
2.
Use the
Focus
button to null the stylus on the sample (use a patterned sample
with easily defined features).
Figure 4.6
Save Zoom Position Dialog Box
Both of the
Zoom positions
should be the same.
Click Save
Zoom Position.
OK
to
save the position.
NOTE:
Side-View optics are P-16+ only.
CAUTION:
Before lowering the head, be sure that the sample is under
the center of the optics, that the stage is not significantly out of level,
and that there are no physical obstacles.
Содержание P-16+
Страница 4: ...KLA Tencor P 16 P 6 User s Guide KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 24: ...KLA Tencor P 16 P 6 User s Guide LOT 4 KLA Tencor Confidential 0142530 000 AB Friday March 13 2009 Preliminary...
Страница 30: ...Introduction Safety Safety KLA Tencor P 16 P 6 User s Guide 1 6 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 180: ...XY View Screen Aligning the Sample KLA Tencor P 16 P 6 User s Guide 4 32 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 196: ...View Scan Window Aborting A Scan KLA Tencor P 16 P 6 User s Guide 5 16 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 434: ...Backup and Restore KLA Tencor P 16 P 6 User s Guide 16 8 KLA Tencor Confidential 0241110 000 AB 3 13 09...
Страница 444: ...KLA TENCOR P 16 P 6 USER S GUIDE REVISION HISTORY 2 KLA TENCOR CONFIDENTIAL 0142530 000 AB 3 13 09...