Scan Recipes - Creating and Editing a Scan Recipe
KLA-Tencor P-16+ / P-6 User’s Guide
3-70
KLA-Tencor Confidential
0142530-000 AB
3/13/09
2D Mean Spacing Sm (1/PC)
3
Mean Peak spacing is the mean value of the local peak spacing of the profile within
the sampling length. The peaks for Peak Count are defined by the
Band
(bandwidth)
parameter from the Peak Count (PC) window.
Spacing
is the inverse of the count.
It is important to note that the
2D Peak Count (PC)
and the
2D Mean Spacing Sm (1/PC)
are related. If running a scan in which these values are to be compared, the
bandwidth
of both must be identical
for the data to have direct correlation.
Use the following procedure to set the 2D Mean Spacing Sm
1.
The option exists to create three 2D High Spot Count parameters. Click in up to
three empty checkboxes to put a check (
9
) in them and activate their variable
boxes.
2.
The bandwidth can be set in either microns (
μ
m) or angstroms (Å). Determine
which units are going to be used and, if a change is necessary, click the menu
arrow under units to display its menu.
3.
Click the desired unit signifier,
μ
m
or
Å
.
4.
The
Band
is the bandwidth bordered equidistant above and below the mean line
of the scan. In most scans, this value is compared to
Peak Count (PC)
so this
Band
(bandwidth) must be identical to the
Band
in
Peak Count (PC)
.
To set or change the
Band
, double-click the current
Band
variable and type in
the new bandwidth.
Histogram Leveling and Depth Analysis
3
This tool has two purposes that are used in the analysis of the gathered data.
First, the leveling of the data is accomplished based on a choice of data to be
used as a leveling basis.
Second, the data is compiled as a histogram for comparison of feature depth in
the scan.
The parameters available in this window work on already accumulated data.
Therefore, the parameters can be adjusted and recalculated over and over again on the
same data to help analyze the scan results.
Both the leveling and the depth analysis histogram are discussed in this section.
Histogram Leveling
3
From the Recipe Editor, click the Histogram Leveling button to open the Histogram
Leveling parameters.
Содержание P-16+
Страница 4: ...KLA Tencor P 16 P 6 User s Guide KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 24: ...KLA Tencor P 16 P 6 User s Guide LOT 4 KLA Tencor Confidential 0142530 000 AB Friday March 13 2009 Preliminary...
Страница 30: ...Introduction Safety Safety KLA Tencor P 16 P 6 User s Guide 1 6 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 180: ...XY View Screen Aligning the Sample KLA Tencor P 16 P 6 User s Guide 4 32 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 196: ...View Scan Window Aborting A Scan KLA Tencor P 16 P 6 User s Guide 5 16 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 434: ...Backup and Restore KLA Tencor P 16 P 6 User s Guide 16 8 KLA Tencor Confidential 0241110 000 AB 3 13 09...
Страница 444: ...KLA TENCOR P 16 P 6 USER S GUIDE REVISION HISTORY 2 KLA TENCOR CONFIDENTIAL 0142530 000 AB 3 13 09...