![KLA Tencor P-16+ Скачать руководство пользователя страница 117](http://html1.mh-extra.com/html/kla-tencor/p-16/p-16_user-manual_1974647117.webp)
0142530-000 AB
KLA-Tencor Confidential
3-69
3/13/09
KLA-Tencor P-16+ / P-6 User’s Guide
Scan Recipes - Creating and Editing a Scan Recipe
4.
The
Height
is the distance
up
from the lowest point of the roughness trace. In
most scans, this value is compared to High Spot Count (HSC) so this height
must be identical to the
Height
in
High Spot Count (HSC)
.
To set or change the
Height
, double-click the current
Height
variable and type in
the new height.
2D Peak Count (PC)
3
Peak Count is the number of peak and valley pairs per unit length projecting through
a band of width
b
centered about the mean line. (See
The Mean line is the line at the mean height of all data. The band is the area bounded
by two lines running parallel to the mean line, at an equal distance from the mean line.
Use the following procedure to set the 2D Peak Count variables.
1.
The option exists to create three 2D Peak Count bandwidth settings. Click in up
to three empty checkboxes to put a check (
9
) in them and activate their variable
2.
The bandwidth can be set in either microns (
μ
m) or angstroms (Å). Determine
which units are going to be used and, if a change is necessary, click the menu
arrow under units to display its menu.
3.
Click the desired unit signifier,
μ
m
or
Å
.
4.
The
Band
is the bandwidth surrounding the mean line. (See
.)
To set or change the
Band
, double-click the current
Band
variable and type in
the new bandwidth. (See
Figure 3.53
Peak Count
1 count
1 count
Mean Line
1 count
b
The boundary lines
are defined by the
band width
entered in
the
Peak Count
Band
variable box.
Figure 3.54
2D Peak Count (PC)
Содержание P-16+
Страница 4: ...KLA Tencor P 16 P 6 User s Guide KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 24: ...KLA Tencor P 16 P 6 User s Guide LOT 4 KLA Tencor Confidential 0142530 000 AB Friday March 13 2009 Preliminary...
Страница 30: ...Introduction Safety Safety KLA Tencor P 16 P 6 User s Guide 1 6 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 180: ...XY View Screen Aligning the Sample KLA Tencor P 16 P 6 User s Guide 4 32 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 196: ...View Scan Window Aborting A Scan KLA Tencor P 16 P 6 User s Guide 5 16 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 434: ...Backup and Restore KLA Tencor P 16 P 6 User s Guide 16 8 KLA Tencor Confidential 0241110 000 AB 3 13 09...
Страница 444: ...KLA TENCOR P 16 P 6 USER S GUIDE REVISION HISTORY 2 KLA TENCOR CONFIDENTIAL 0142530 000 AB 3 13 09...