3-Component Force Link, Type 9327C
Page 34
9327C_002-477e-05.10
9.2
Dimensions 3-Component Force Link Type 9327C
Fig. 15: Dimensions 3-component force link Type
9327C
Страница 1: ...Instruction Manual 3 Component Force Link Type 9327C 9327C_002 477e 05 10...
Страница 2: ...Instruction Manual 3 Component Force Link Type 9327C 9327C_002 477e 05 10 Fz Fx Fy...
Страница 3: ...o change without notice Kistler reserves the right to change or improve its products and make changes in the content without obliga tion to notify any person or organization of such changes or improve...
Страница 4: ...sing a Dynamometer 13 5 2 6 Component Force Measurement Using a Dynamometer 15 5 3 3 Component Force Measurement Using a Single Force Sensor 16 5 4 Measuring System with 3 Component Force Link 21 5 5...
Страница 5: ...Content 9327C_002 477e 05 10 Page 3 10 Annex 38 10 1 Glossary 38 10 2 Measurement Uncertainty 41 10 3 Linearity 42 10 4 Frequency Range 44 10 5 Influence of Temperature 45 10 6 Index 47 Total pages 47...
Страница 6: ...tion and acoustic emission Strain gage sensor systems for measuring force and tor que Piezoresistive pressure sensors and transmitters Signal conditioners indicators and calibrators Electronic control...
Страница 7: ...erating instructions The sensor must be installed operated and maintained only by persons who are familiar with the instrument and are adequately qualified for the work concerned If it is accepted tha...
Страница 8: ...than 1013 In order to preserve this value all plug and socket connections must be kept absolutely clean and dry The insulation resistance can be meas ured with the insulation tester Type 5493 Protect...
Страница 9: ...access to the information you require Please keep these instructions in a safe place where they are always to hand If you lose these instructions please contact your Kistler customer service organiza...
Страница 10: ...ree orthogonal compo nents of any force acting on the sensor Fx Fy and Fz 3 component force links measure Cutting forces during machining Impact forces in crash tests Recoil forces of rocket engines V...
Страница 11: ...on of the Instrument 9327C_002 477e 05 10 Page 9 Inertia Force Measurement Fig 2 Force limited vibration test Product Testing Fig 3 Elastomer test stand Reaction Force Measurement Fig 4 Rocket motor s...
Страница 12: ...plates of which one is sensitive to pressure in the z direction and the two others to shear forces in the x and y directions Mea surement takes place practically without displacement The force introdu...
Страница 13: ...component The latter has to be fixed firmly to the force link top plate The force link can be used as a single measuring ele ment or several of them with a common top plate force plate 4 2 Mounting of...
Страница 14: ...e 12 9327C_002 477e 05 10 Mounting the top plate with 4 each M6 screws through the top plate Recommended Tightening Torques for Fixing Screws Thread Property Class ISO898 1 Tightening Torques MA N m M...
Страница 15: ...g nals each of Fx Fy and Fz of the four sensors are summed However the sum of the four sensors always shows the correct value irrespective of the force application point A 3 component dynamometer meas...
Страница 16: ...3 Component Force Link Type 9327C Page 14 9327C_002 477e 05 10 Fig 6 Force plate with solid top plate Fig 7 Force plate with top plate of any shape...
Страница 17: ...measurement the output signals of the four 3 component force sensors are interconnected as in Fig 8 Fig 8 Interconnection of the twelve sensor outputs fort he 6 component force measurement The Three...
Страница 18: ...an individual preloaded 3 component force sensor can only be utilized when the line of action of the resulting force vector runs through the center of the sensor i e the origin of the coordinate syste...
Страница 19: ...point and the three compo nents of the resulting force are known the active moments can be calculated as follows Mx Fz ay Fy az My Fx az Fz ax Mz Fy ax Fx ay When calculating the moments it is essent...
Страница 20: ...Preload force Fv 20 kN Coefficient of friction H 0 2 Force application point see Fig 10 Coordinates ax 4 mm ay 3 mm az 25 mm Process forces Fx 215 N Fy 130 N Fz 2 160 N Moments Mx Fz ay Fy az 9 7 N m...
Страница 21: ...Measuring Range Introduction of Force 9327C_002 477e 05 10 Page 19 Fig 11 Nomogram to determine the max possible moment load Example...
Страница 22: ...9327C_002 477e 05 10 Result max possible bending moment Mb My 2 Mx 18 N m max possible moment Mz 20 N m The sensor load is within the specifications The bending moment load is approximately 94 4 the m...
Страница 23: ...3 Component Force Link 5 5 Measuring System with four 3 Component Force Links Dynamometer 1 see data sheet cables for multi component force sensors dynamometers and force plates 1687B_000 545 2 see d...
Страница 24: ...all Force Changes The piezoelectric method of measurement allows very accurate measurement of small force changes when there is large preload present at the same time When for exam ple fluctuations Fz...
Страница 25: ...1 N V for Fz f Set the amplifier to Short Time Constant g Switch the amplifier to Operate The Fz variations can now be measured with great accuracy since the dy namic response threshold is around 0 02...
Страница 26: ...a drift up to 10 mN s can occur Depending on the measuring range therefore measure ments from a few minutes up to several hours are possible The illustration shows the example of a frequency curve in...
Страница 27: ...ntroduction parts the higher is the resonant frequency and the wider is the usable fre quency range At the same time the mass mounted on the force link must be kept low Most piezoelectric force sensor...
Страница 28: ...sis of sensor Repeatability of sensor Error of charge amplifier Average Linearity of charge amplifier The error can be improved if calibration is done in a partial range Electrical Drift The piezoelec...
Страница 29: ...force link by tests Temperature Gradient Error The temporarily occurring maximum signal output of a sensor exposed to a temperature change is defined as Temperature gradient error The temperature gra...
Страница 30: ...he error of the amplifier depends on the Type and the range capacitor It can vary between 0 5 FSO 2 FSO for lab Type charge amplifiers and 1 FSO 5 FSO for industrial charge amplifiers To improve the e...
Страница 31: ...t the output signal stays between 5 10 V For normal force measurement set low pass filter to 1 3 of the resonance frequency Set charge amplifier to operating mode immediately before force measurement...
Страница 32: ...e calibrations can even be done by our subsidiaries either by identical or by similar procedures Kistler keeps complete records available to our customers and allowing the calibration and repair recor...
Страница 33: ...sor are machined parallel and flat within very close tolerances These surfaces should not be levelled with a whetstone Clean plug connectors of the sensor or connecting cable regularly with a cleaning...
Страница 34: ...nsulation tester Type 5493 Clean plug connec tions During the measurement the charge amplifier goes into overload The force to be measured is too high for the measuring range selected Select a higher...
Страница 35: ...Force application point 8 mm below the surface of the cover plate Fx Fy kN 0 1 0 0 1 Calibrated Range Force application point centric Fz kN 0 8 0 0 8 Permissible moment load Mz 0 Fz 0 Mx My N m 22 22...
Страница 36: ...3 Component Force Link Type 9327C Page 34 9327C_002 477e 05 10 9 2 Dimensions 3 Component Force Link Type 9327C Fig 15 Dimensions 3 component force link Type 9327C...
Страница 37: ...ories The following accessories can be obtained for sensor Type 9327C Fig 16 Connecting cable Type 1698AA see data sheet 16xx_000 545 Fig 17 Connecting cable Type 1698AB see data sheet 16xx_000 545 Fi...
Страница 38: ...the dynamometer A maximum of eight measuring signals is available at the output socket similar to the Kistler standard dynamome ters Connection to the charge amplifier is made with a 3 core cable for...
Страница 39: ...f summing box Type 5417 Fig 22 Cleaning and insulation spray 250 ml Type 1003 Fig 23 Insulation tester Type 5493 Ordering Code Type 3 Component Force Link 9327C 42x42x42 mm 8 8 kN Input Fx1 Fy1 Fz1 Fx...
Страница 40: ...ts of the factory calibration Charge amplifier Part of a measuring chain which converts the charge signal from the sensor into a proportional voltage signal Charge signal Output signal in picocoulomb...
Страница 41: ...tic which is measured e g force torque etc Measuring chain Interconnection of several individual components to meet measuring requirements Measuring chains usually consist of sensors and amplifiers in...
Страница 42: ...tem which produces a definite change in the output signal as a function of the change of the measurand acting on it Temperature influence see Annex 10 5 Threshold Smallest change in the measurand whic...
Страница 43: ...tion of the sensor and environmental influ ences For the highest accuracy requirements we recom mend calibration of measuring ranges specific to the appli cation Measurement uncertainty of charge ampl...
Страница 44: ...ain unavoid able deviations occur due to the mechanical construction of the sensor Linearity represents the maximum deviation be tween ideal and actual output signal characteristics in rela tion to th...
Страница 45: ...res 0 Linearity a resmax resmin Fig 24 Best straight line linearity and hysteresis Hysteresis Quartz itself has a scarcely measurable hysteresis How ever the mechanical construction of the sensor can...
Страница 46: ...sing resonance rise Key f Measuring frequency fn Natural frequency A A0 Amplitude ratio The following approximate values apply to the amplitude error or achievable accuracy as a function of frequency...
Страница 47: ...s then con tinuously varied until its natural frequency coincides with that of the actual sensor Technical data An equivalent body defined in this way is usually a good approximation of the sensor The...
Страница 48: ...perature con stant during the actual measuring time Fig 26 Example of a temperature error of the output signal zero for a temperature rise from 150 C to 200 C immersion bath Temperature coefficient of...
Страница 49: ...force measurement 9 Insulation resistance 6 40 Isolationspr fger t 37 L Linearity 40 M Measurand 40 Measuring chain 40 Measuring range 40 Measuring uncertainty 42 Moments 18 N Natural frequency 40 45...