Trouble. When there is in case of a continuous " Check CIS " message
Solution 1. Remove dust on the CIS.
Solution 2. Update the FPGA SW again using USB
Solution 3. Update the APL SW again using USB
Solution 4. Update the DSP SW again using USB
Trouble. When alert window does not disappear after removal of JAM
Solution 1. Try again open/close Main Cover & Back Cover
Solution 2. Update the Mecha SW again using USB
Solution 3. Replace the IR sensor PCB
Simple Trouble shooting guide 2
19
Содержание Newton mini
Страница 1: ...Newton mini USER MANUAL ...
Страница 25: ...3 Stack Sensor Clean the sensor as shown in the photo with a soft towel or a brush 25 ...
Страница 26: ...4 The dust pocket Pull out a dust tray and clean 26 ...
Страница 27: ...Thank you KISAN will make continuous efforts to improve our products ...