Tab. 4: Interfaces
Interface
Module
LAN interface
PC-KIT-NET (optional)
USB port
PC-KIT-STICK
Potential-free alarm contact
Remote warning system (for
example GSM-MODUL or con-
nection to building control
system (see circuit diagram on
unit))
For information about connecting the modules to the unit, see the
relevant product documentation.
Connect only compatible USB flash drives!
Connect only modules with the following properties
to the USB port:
–
Maximum memory capacity of 32 GB
–
Formatted in FAT-32 format
Remove USB flash drive after use!
The USB flash drive must not remain permanently
in the unit.
–
Remove USB flash drive after the data transfer
“Error messages and status dis-
plays of the PC-KIT-STICK” on page 72.
1.5 Unit functions
1.5.1 Cooling
The unit is equipped with a cooling machine with interior evapo-
rator for cooling chilled goods.
After initial commissioning or recommissioning, the cooling
machine takes a while to cool the interior down to the set target
temperature value.
With circulation cooling, the air circulates around the interior of the
refrigerator. Circulation cooling is performed by means of ventila-
tion plates and a fan.
Circulation cooling reduces physically induced temperature differ-
ences and the target temperature is kept constant throughout the
interior.
Cooling machine
Circulation cooling
Product description
Unit functions > Cooling
26.05.2021
Refrigerator/freezer BL series
15