Intended Use
The forehead thermometer is designed for body surface and forehead temperature
measurement for infant and adults without contact to the human body.
The forehead thermometer can also be used to measure the
temperature of a baby’s bottle,
bath and room temperature (by using the surface mode).
Normal Temperatures According to Measurement Method:
Measurement Method
Normal Temp Range ºC
Normal Temp Range ºF
Rectal
36.6 to 38
97.8 to 100.4
Oral
35.5 to 37.5
95.9 to 99.5
Axillary
34.7 to 37.3
94.4 to 99.1
Ear
35.8 to 38
96.4 to 100.4
The temperature of the human body can vary throughout the day. It can also be influenced
by numerous external factors.
Normal Temperature According to Age
Age Method
Normal Temp Range ºC
Normal Temp Range ºF
0
–
2 years
36.4 to 38.0
97.5 to 100.4
3
–
10 years
36.1 to 37.8
97.0 to 100.0
11
–
65 years
35.9 to 37.6
96.6 to 99.7
>65 years
35.8 to 37.5
96.4 to 99.5
Attention: ASTM laboratory accuracy requirements in the display of 37 to 39
℃
(98 to 102
℉
)
for Skin IR thermometers is ±0.3
℃
(± 0.5
℉
)
,
whereas for mercury in-glass and electronic
thermometers, the requirement per ASTM Standards E667-86 and 1112-86 is ±0.1
℃
(± 0.2
℉
)
Manufacturer will make available upon request specific instructions for tests to determine
the laboratory error, clinical bias and clinical repeatability.
When you need to clinical accuracy characteristics and procedures, please contact with the
local distributors obtain from our company