7-36
TOS9213S/ TOS9213AS
7.4.5
System-related messages
This section describes the messages to be set on the system screen.
MEASMODE (MMOD)
Sets the display mode (MEAS MODE) of the measured current (or measured resis-
tance). Also inquires about the display mode of the measured current (or measured
resistance).
Message
Setting:
MEASMODE_<{NORM|MAX}>
MMOD_<{NORM|MAX}>
Query:
MEASMODE?
MMOD?
Program data
Data format: Character
Set value:
NORM: Real Time; MAX: Max Hold (Min Hold)
(Example) To set the display mode of the measured current (or resistance) to Max
Hold (Min Hold for resistance),
MMOD MAX
Response
To MMOD
?
, the present display mode of the measured current (or resistance) is
returned as 0 for NORM and 1 for MAX.
(Example) If the display mode of the present measured current is set to Real Time,
0
is returned.
PASSHOLD (PHOL)
Sets the PASS hold time in a PASS judgment. Also inquires about the current PASS
hold time.
Message
Setting:
PASSHOLD_<
set value
>
PHOL_<
set value
>
Query:
PASSHOLD?
PHOL?
Program data
Data format: Real number, character
Set value:
0.2 to 10.0, HOLD
Unit:
s
(Example) To set the PASS hold time to HOLD,
PHOL HOLD
Response
To PHOL
?
, the current PASS hold time is returned.
(Example) If the current PASS hold time is 0.5 s,
0.5
is returned.