12 General
TOS8030
■
Displaying the cause of activation of the protective function as
a code number
If the protective function is activated, the cause is given as a code number via the
voltmeter. A code number is also indicated if there are discrepancies in setting test
conditions, allowing you to promptly correct the setting based on the code number
displayed.
■
Remote control
The optional remote control box or test probe allows testing to be started or stopped
remotely.
■
Status signal output
The tester has output terminals for READY, H.V ON, PASS, FAIL, and PROTEC-
TION signals to enable external monitoring of tester status.
Use this function with the remote control function to automate functions or to
reduce actual hands-on testing requirements.
■
Sequence circuit with noise reduction features
For reliability, the sequence circuit incorporates thorough noise reduction features
to prevent noise-induced malfunctions.
1.3
Options
Remote Control Boxes
Connecting a remote control box to the REMOTE connector of our Withstanding
Voltage Testers, Insulation Resistance Testers, or Withstanding Voltage Insulation
Resistance Testers, enables remote starting or stopping of testing.
The RC01-TOS has one START switch; the RC02-TOS has two START switches.
With the RC02-TOS remote control box, testing can be started only when both
START switches are pressed simultaneously.
Description of RC01-TOS/RC02-TOS
OPERATE
switch
The START switch is enabled only when this switch is ON. Turning this
OFF halts testing.
START
switch
When the OPERATE switch is ON and the tester is in a READY state, press
this switch to begin testing.
STOP
switch
This switch shuts off the output voltage or cancels a status (such as FAIL). It
has the same function as the STOP switch on the tester.
RC01-TOS
200(W) x 70(H) x 39(D) mm
RC02-TOS
330(W) x 70(H) x 39(D) mm