SYSTem Command
KIKUSUI Electronics Corp.
PWR-01 Interface Manual
SYST:CONF:STAR:PRI
Sets the operation mode to be prioritized when the output is turned on.
This command is invalid while a sequence is running.
Setting the slave unit during master-slave parallel operation is invalid.
Command
SYSTem:CONFigure:STARtup:PRIority <character>
SYSTem:CONFigure:STARtup:PRIority?
Parameter
Value:
CC
CC (constant current) is prioritized.
CV
CV (constant voltage) is prioritized.
(Example)
SYST:CONF:STAR:PRI CC
Response
Returns the state that the PWR-01 starts up in when the output state at power-on is
set to on in character format in response to SYST:CONF:STAR:PRI?.
SYST:ERR
Reads the oldest error information or event information from the error queue. The
error queue can store up to 16 errors.
-> Tutorial “Error Checking”
Use the *CLS command to clear the error queue.
Command
SYSTem:ERRor[:NEXT]?
Response
Returns the oldest error or event information in the error/event queue in response
to SYST:ERR? as follows:
(Example) When there are no errors or events
+0"No error"
(Example) When a command that cannot be executed under the current operating
conditions has been received
-221,"Settings conflict"