PMX_MULTI
User’s Manual
91
Specifications
Parallel operation and series operation
PMX32-3DU
PMX32-3TR
PMX32-2QU
Parallel operation
Applicable channels
Master: CH2, slave: CH1
Constant
voltage
Operating range
0 V to 32 V
Setting range
0 V to 33.6 V
Setting accuracy
*1
*1.
At an ambient temperature of 23 °C±5 °C.
0.3 % set + 0.1 % rating
Resolution
1 mV
constant
current
Operating range
0 A to 6 A
0 A to 4 A
Setting range
0 A to 6.3 A
0 A to 4.2 A
Setting accuracy
0.4 % set + 0.1 % rating
Resolution
0.2 mA
Voltmeter
Maximum
display
99.999 (fixed decimal point)
Display
accuracy
±(0.5 % of r 10 digit)
Ammeter
Maximum
display
9.999 (fixed decimal point)
Display
accuracy
±(1 % of r 10 digit)
Series operation
Applicable channels
Master: CH2, slave: CH1
Constant
voltage
Operating range
0 V to 64 V
Setting range
0 V to 67.2 V
Setting accuracy
*1, *2
*2.
The value is measured at the sensing point.
0.3 % set + 0.1 % rating
Resolution
2 mV
constant
current
Operating range
0 A to 3 A
0 A to 2 A
Setting range
0 A to 3.15 A
0 A to 2.1 A
Setting accuracy
0.4 % set + 0.1 % rating
Resolution
0.1 mA
Voltmeter
Maximum
display
99.999 (fixed decimal point)
Display
accuracy
±(0.5 % of r 20 digit)
Ammeter
Maximum
display
9.999 (fixed decimal point)
Display
accuracy
±(1 % of r 5 digit)