PLZ-4W
Applied Operation 6-27
6.6.4
Overview of the Fast Sequence
To create a sequence, you must understand the concept of “program” and “step.”
Fig. 6-16
Conceptual diagram of program and step
• The operation modes that can be used in a fast sequence are CC and CR modes. In
the explanation below, CC mode is used. When using CR mode, read the current
values as resistance values.
•
Program
The fast sequence can be saved only to program 11. Set the operation mode to CC or
CR. The only method of executing the program is looping. Programs cannot be
chained as in the normal sequence. The number of programs that can be saved is 1
(Program 11).
Below are the settings of a program.
• No.
Program number (11)
• Memo
Memo (up to 11 characters can be specified)
• Mode
Operation mode of the program (CC or CR)
• Range
Range of the program (Current range - Voltage range)
• Loop
Number of program loops (1 to 9999 where 9 999 is infinite
loop)
• Last Load
Load condition (OFF/ON) at program end
• Last Set
Current value at program end (0 to 100 % of the range setting)
• RPTSTEP
Last step number (3 to 1024) per loop
• TIME BASE
Step execution time (25
s to 100 ms)
Resolution
25
s to 0.1 ms
25
s
0.1 ms to 100 ms
100
s
Step 0001
Step 0002
Step 0003
Step 1024
Program 11
Maximum number
of steps is 1024
s
s
s
NOTE