background image

PLZ-5W SR Series Setup Guide

11

1.Preparation

    Operating the Load Input Terminal Cover

Closing the load input terminal cover

1

Close the load input terminal cover.

2

Tighten the screws holding the load input terminal cover on the rear 

panel.

Fasten the covers correctly to prevent them from moving.

Screw

Load input terminal cover

Содержание PLZ-5W SR Series

Страница 1: ...s 5 Preparation 7 Connecting the Power Cord 7 Checking Whether the Power Is On or Off 8 Operating the Load Input Terminal Cover 10 Connecting the EUT 13 Connecting the Load Cables 13 Connecting the Se...

Страница 2: ...nce briefly explains the control panel and the basic operation of it Setup Guide this guide This guide is intended for users who are using the PLZ 5W SR series for the first time It covers topics rela...

Страница 3: ...ntially hazardous situation which if ignored may result in damage to the product or other property Indicates information that you should know Load the included CD ROM into the CD ROM drive After a mom...

Страница 4: ...e product in place by lowering the stoppers to the floor When moving this product be sure to observe the Pre cautions to Be Taken When Moving the Product in the Safety information manual The following...

Страница 5: ...005W SR 8 units 3 Front panel load input terminal cover Not used Do not remove the cover 4 Display Displays the settings measured values and other information 5 Handle Handle for moving the product on...

Страница 6: ...connector External control connector p 20 5 DC OUT 5V connector Used with the optional GPIB converter PIA5100 For details see the GPIB con verter manual contained in the included CD ROM PIA5100 Operat...

Страница 7: ...ectric shock The product is grounded through the power cord ground wire Connect the protec tive conductor terminal to earth ground Use the included power cord to connect to the AC line If the supplied...

Страница 8: ...the load value input screen appears If you notice strange sounds unusual odors fire or smoke around or from inside the PLZ 5W SR turn off the power switch or turn off the switchboard breaker By facto...

Страница 9: ...medy Nothing is shown on the display Check that the power cord is not broken and that the connection at the AC INPUT terminal is secure An alarm is occurring See Alarm types and operation and Clearing...

Страница 10: ...r making the connections be sure to close the cover Adjusting the position of the lock plate provided with the load input terminal cover makes the opening narrower for safer use Opening the load input...

Страница 11: ...d Input Terminal Cover Closing the load input terminal cover 1 Close the load input terminal cover 2 Tighten the screws holding the load input terminal cover on the rear panel Fasten the covers correc...

Страница 12: ...ening according to the load cables and other various cables in use Factory default condition Lock plates 5 pcs Screw M3 Remove Move the lock plate to adjust the position Example When four flexible con...

Страница 13: ...ecting the load cables Select load cables that are suitable for the current to be used You can use the optional large current load cables For details on large current load cables see the large current...

Страница 14: ...the cables as much as possible to increase heat radiation enables a greater amount of current to flow However running the positive and negative output load cables side by side or bundling them togethe...

Страница 15: ...minal cover p 11 WARNING Risk of electric shock Do not touch load input terminals when the output is turned on The load input terminals on the front panel are connected internally to the load input te...

Страница 16: ...inal cover p 10 2 Connect the sensing cables Remove the two short bars from the sensing terminals and then connect the cables Short bars must be connected to the sensing terminals when remote sensing...

Страница 17: ...g Cables 5 Connect the sensing cables to the EUT Connect the cables as shown in the figure Connect the cables as close to the EUT as possible For details on enabling the remote sensing function see th...

Страница 18: ...your Kikusui agent or distributor 1 Turn the POWER switches of all the PLZ 5W SRs and PLZ 5W series units that you will connect 2 Open the load input terminal cover p 10 3 Connect a LAN cable to the O...

Страница 19: ...s Connection for Synchronized Operation 7 Connect all PLZ 5W SR and PLZ 5W series units with LAN cables Connection example when connecting the PLZ15005W SR PLZ205W PLZ1205W This completes the connecti...

Страница 20: ...ity formerly AMP catalog For information about how to obtain these tools or replacement parts contact your Kikusui agent or distributor Connecting to the EXT CONT connector 1 Open the load input termi...

Страница 21: ...nector on the rear panel For details on remote control and how to use it see the Communication Interface Manual on the included CD ROM 1 Open the load input terminal cover p 10 2 Connect a cable to th...

Страница 22: ...Copyrights The contents of this manual may not be reproduced in whole or in part without the prior consent of the copyright holder The specifications of this product and the contents of this manual a...

Отзывы: