88
PFX2532
Preparing to Connect the DUT (Battery)
Cables
1
for connecting the OP02-PFX and the input terminal to the DUT (battery) are not included.
Prepare the following cables.
■
Cables and tools that are necessary to make the connection
1
The TL09-PFX option consists of a dedicated voltage sensing cable and thermocouple for
connecting the DUT (battery) to a PFX2532 in which an OP02-PFX is installed. (
See
Cable
DUT cable
Cable for running charge/discharge current between the PFX2532
and DUT (battery)
Voltage sensing cable
Cable for sensing the electrode voltage of the DUT (battery)
Temperature sensing cable
Cable for sensing the temperature of the DUT (battery)
Thermocouple (type K)
Element-wire diameter: 0.20 mm
2
or more
Cell voltage sensing
cable
Cable for sensing the cell electrode voltage of the DUT (battery)
cable 0.20 mm
2
(AWG24) to 0.52 mm
2
(AWG20), shielded cable
Flat-blade screwdriver
Core diameter: Ø3, tip width: 2.6 mm
Wire stripper
A wire stripper that matches the wires listed above
See
Example of an OP02-PFX Volt/Thermometer Unit installed in slot 1;
only the wiring between the PFX2532 and the DUT (battery) is shown
+
–
+
–
+
–
+
–
+
–
+
S
–
S
+
T
–
T
OP02-PFX
DUT
cable
Thermocouple (5 m max.)
Cell voltage
Sensing
cable
(shielded cables)
Thermistor
Cell V Term1
Cell V Term2
Cell V Term3
Cell V Term4
Cell T Term1
Cell T Term2
Cell T Term3
Cell T Term4
Voltage sensing
cable (shielded cables)
Temperature
sensing cable
(shielded cables)
PFX2532
The DUT (battery)
Содержание PFX2532
Страница 12: ...12 PFX2532 This page is intentionally blank ...
Страница 25: ...Connecting and Configuring Devices This chapter describes how to connect and configure the devices ...
Страница 62: ...62 PFX2532 This page is intentionally blank ...
Страница 71: ...Specifications This chapter provides the PFX2532 s specifications and outline drawing ...
Страница 96: ...96 PFX2532 This page is intentionally blank ...
Страница 141: ......