PFX2532
81
PFX2532 Electric Specifications
5
Specific
a
tions
Temperature measurement
Thermistor 103AT-2 (by SEMITEC Corporation) is used for the temperature detector.
Measurement accuracy (continued)
High-speed sampling
Current
measurement
Range
8
-200.000 A to 200.000 A
Accuracy
1 ms sampling
±(0.4 % of r 0.5 % of rating)
10 ms sampling
±(0.3 % of r 0.1 % of rating)
100 ms sampling ±(0.2 % of r 0.1 % of rating)
Resolution
1 mA. The same for all sampling settings.
Voltage
measurement
Range
60 V range
-6.0000 V to 60.0000 V
3
6 V range
-1.0000 V to 6.0000 V
4
Accuracy
1 ms sampling
5
±(0.1 % of r 0.1 % of rating)
10 ms sampling
5
±(0.1 % of r 0.05 % of rating)
100 ms sampling 60 V range: ±(0.05 % of r 0.02 % of rating)
6 V range: ±(0.05 % of r 0.04 % of rating)
Resolution
0.1 mV. The same for all sampling settings.
1
Measurable range: -210.000 A to 210.000 A (TYP). However, the accuracy is not guaranteed for values outside of
the range listed in the table.
2
Ambient temperature at 18
°
C
to 28
°
C.
3
Measurable range: -6.500 V to 65.000 V (TYP). However, the accuracy is not guaranteed for values outside of the
range listed in the table.
4
Measurable range: -6.500 V to 6.500 V (TYP). However, the accuracy is not guaranteed for values outside of the
range listed in the table.
5
The same for the 6 V range and 60 V range.
6
Accuracy of the elapsed time (Cutoff condition) when charging / discharging or resting.
7
Monthly error: approximately 30 seconds.
8
The accuracy is not guaranteed for values outside of the rated output range.
9
The wavering caused by the ripple noise and AC line noise (50 Hz/ 60 Hz) of the DC power supply that you are
using is not included.
Resistance (temperature) measurement section
1
Measurable range
-40.0 °C to 100.0 °C
Measurement resolution
0.1 °C
Measurement accuracy
2
3
±0.5 °C (Measurement temperature at 0 °Cto 40.0 °C)
±1 °C (Measurement temperature at -20 °C to 80 °C)
Reference (thermistor 103AT)
Model
103AT-2 by SEMITEC Corporation
R25
Nominal zero-load resistance at 10.0 kΩ, 25 °C
Operating temperature range
-50.0 °C to 110.0 °C
Temperature accuracy
±0.5 °C (Measurement temperature at 0 °C to 40.0 °C)
Tolerance
±1 %
Constant B
3435K±1 % (Measurement temperature at 25 °C)
1
The temperature measurement does not trace the absolute temperature. Temperature converted from
resistance.
2
Excludes temperature detector errors.
3
Ambient temperature at 18
°
C
to 28
°
C.
Содержание PFX2532
Страница 12: ...12 PFX2532 This page is intentionally blank ...
Страница 25: ...Connecting and Configuring Devices This chapter describes how to connect and configure the devices ...
Страница 62: ...62 PFX2532 This page is intentionally blank ...
Страница 71: ...Specifications This chapter provides the PFX2532 s specifications and outline drawing ...
Страница 96: ...96 PFX2532 This page is intentionally blank ...
Страница 141: ......