Troubleshooting
This appendix explains troubleshooting measures for typical problems. Check whether or not your
problem corresponds to any of the items listed below. In some cases, the problem can be solved
easily.
If your problem does not correspond to any of the listed items, If you carry out the corrective action
but the situation does not improve, contact your Kikusui agetnt or distributor.
Symptom concerning Charge / Discharge System Controller PFX2512
The power does not turn on.
Unable to perform testing properly.
Problem
Items to check and corrective actions
The "STANDBY" LED does not
light on even when the POWER
switch is turned on.
Is the power cable connected properly?
Are those cables between the equipments connected properly?
When the POWER switch is
turned on, the "ALARM/
WARNING" LED Illuminates.
The "ALARM/WARNING" LED may illuminate in red for rare case. However, it
does not indicate as an abnormal condition, it is because of the detection of
the momentary interruption activates when turning on the POWER switch.
Please turn off the POWER switch, then wait for more than 5 seconds and turn
on the POWER switch again.
ー
Is any of protection functions or alarm functions activated?
See
Problem
Items to check and corrective actions
When it is in the "Idle" state, the
"ALARM/WARNING" LED
Illuminates in orange color.
When it is in the "Idle" state even if the charge/discharge test is not being
executed, the alarm function activates. Please remove the cause of occurrence.
Test Condition Editor, Test Executive
After the test is started, the
"ALARM/WARNING" LED
Illuminates in red color.
The test is interrupted because of an abnormal operation was detected, and it
becomes in the state of alarm. Please remove the cause of occurrence.
Test Condition Editor, Test Executive
The channel is not identified and
the "STANDBY" LED Illuminates.
Is the setting channel number on the rear panel of the PFX2512 series same for
other PFX2512's ?
The PC is suddenly shut down
when it is used.
Is the PC in the state of the standby or hybernation mode? Please make sure to
turn off the setting for shifting the state to the standby and or hybernation mode.
-
When the PC is shutting down or freezing while the execution of the charge
and discharge test, ongoing test of the charge or discharge contiues for its
operation, however, the test data is not subject to be saved. To restart the test,
please refer to the Help file of the BPChecker3000.
Test Executive
The test can not be executed.
Is the model ID set correctly?
Does the "ALARM/WARNING" LED (Orange) illuminate?
The vibration sensor may have been activated.
Confirm that the PFX2512 is located on the stable surface in horizontal
position.
See
?
Help
?
Help
?
Help
Содержание PFX2515
Страница 10: ...10 PFX2512_CE This page is intentionally blank ...
Страница 11: ...General Description This chapter describes the outline of product the connectable equipments and the options ...
Страница 58: ...58 PFX2512_CE This page is intentionally blank ...
Страница 59: ...Specification This chapter contains the PFX2512 specifications and outline drawings ...
Страница 86: ...86 PFX2512_CE This page is intentionally blank ...
Страница 100: ...100 PFX2512_CE This page is intentionally blank ...
Страница 103: ......