1-20 Setup
PFX2000
Fig. 1-11
TP-BUS Connection
• Connect unit frames with frame addresses 1 to 6 to TP-BUS 1; connect unit
frames with frame addresses 7 to C to the TP-BUS 2.
• If you are using the impedance measurement unit, be sure to connect it to TP-BUS
1.
MADE IN JAPAN
TP--BUS
USB
AC LINE
POWER
MADE IN JAPAN
FUSE
TERMN
S1
TP-BUS
ON
KIKUSUI ELECTRONICS CORP.
FRAME 1
FRAME 2
FRAME 3
FRAME 4
FRAME 5
FRAME 6
FRAME 7
FRAME 8
FRAME 9
FRAME 10 FRAME 11 FRAME 12
CONT
DUT
VA
MAX
25
FREQUENCY
50
/
60Hz
LINE VOLTAGE
RANGE
100 - 240V
FUSE
1A (T)
AC250V
No.
FRAME
DUT
CONT
TERMN
TP-BUS
TRIP
+
-
BATTERY TESTER FRAME UNIT
No.
KIKUSUI ELECTRONICS CORP.
OFF
ON
MADE IN JAPAN
1
2
3
4
5
6
7
8
9
10
AC LINE
IMPEDANCE
VA MAX
1.5k
FREQUENCY
50
/
60 Hz
LINE VOLTAGE RANGE
100-120V
200-240V
IF THIS COLUMN IS BLANK ,
THE UNIT IS WIRED IN 200V
SET TING SUPPLY
STANDARD
OUTPUT
TERMINAL
No. 1- 10
0
E
D
C
B
A
9
8
7
6
5
4
3
2 1
— +
— -
—+S
—-S
—+ T
—- T
—FG
Impedance
Measurement Unit
Control unit
Unit frame
No. 1
Core
Core
Core
TERMN switch
ON
TERMN switch
OFF
TP-BUS 1 (twisted-pair cable)
Total length: 200 m or less
(between the devices at each end on the bus)
Number of twists: 1 turn/cm
TP-BUS 2
Unit frame
No. x
(can connect up to 6 units)
NOTE