256
SYST:CONF:PHAS:UV
Sets the U-V phase difference.
This command is valid for three-phase output.
If the signal source is set to external signal source (VOLT:EXT:EXTDC:SIGN:SOUR
EXT), this command is invalid.
Command
SYSTem:CONFigure:PHASe:UV <numeric>
SYSTem:CONFigure:PHASe:UV?
Parameter
Value:
-360.00 to 360.00 U-V phase difference
The default value for three-phase output is 120.
Unit:
DEG
Settings are reset to default values when an *RST command is sent.
Example
SYST:CONF:PHAS:UV 121.0
Response: NR3