Protection Features and Alarms
2
Ba
sic Featu
res
State Display of the Limit Features
When the voltage V-LIMIT protection, the current I-LIMIT protection, or the power limit
(internal loss) protection is activated, the limit state display shows which protection feature
has been activated.
How to Set and Check Activation Points
Mode and selectable items
Activation points
Protection feature
Display
Voltage limit
V-LIMIT
+V.LIM: The voltage was limited by the positive voltage activation point.
-V.LIM
1
: The voltage was limited by the negative voltage activation point.
1
In unipolar mode, -V.LIM is ignored.
Current limit
I-LIMIT
+I.LIM: The current was limited by the positive current activation point.
-I.LIM: The current was limited by the negative current activation point.
Power limit
P-LIMIT
P.LIM: The power was limited by the sink power activation point.
An example where all the limit features are displayed
Mode
CURR and
VOLT
(selection)
OVP/OCP ACTION
(selection)
OVP, OCP, and
LIMIT
(numerical setting)
Overvoltage
protection
OVP (OUTPUT-OFF)
OVP
OUT OFF
+OVP
-OVP
OVP (POWER-OFF)
POW OFF
V-LIMIT
LIMIT
None
+V.LIM
-V.LIM
Overcurrent
protection
OCP (OUTPUT-OFF)
OCP
OUT OFF
+OCP
-OCP
OCP (POWER-OFF)
POW OFF
I-LIMIT
LIMIT
None
+I.LIM
-I.LIM
Screen
page
Setting
Description
Settable range (resolution)
1
1
Underlined values are factory default settings.
PBZ20-20
PBZ40-10
(1/2)
+I.LIM
+OCP
Positive (+).
Current activation point.
+0.2 A to +22.0 A
(0.1 A)
+0.1 A to +11.0 A
(0.1 A)
-I.LIM
-OCP
Negative (-).
Current activation point.
-0.2 A to -22.0 A
(0.1 A)
-0.1 A to -11.0 A
(0.1 A)
+V.LIM
+OVP
Positive (+).
Voltage activation point.
+0.2 V to +22.0 V
(0.1 V)
+0.4 V to +44.0 V
(0.1 V)
-V.LIM
2
2
Cannot be set when the PBZ is in unipolar mode.
-OVP
2
Negative (-).
Voltage activation point.
-0.2 V to -22.0 V
(0.1 V)
-0.4 V to -44.0 V
(0.1 V)
(2/2)
CURR
Overcurrent protection
feature.
LIMIT or OCP
VOLT
Overvoltage protection
feature.
LIMIT or OVP
OVP/OCP ACTION
The protective action that
is carried out when OVP
or OCP is activated.
OUT OFF or POW OFF
Содержание PBZ20-20
Страница 11: ...PBZ 11 ...
Страница 18: ...18 PBZ ...
Страница 75: ...Memory Response and Configuration This chapter explains preset memory setup memory response and configura tion ...
Страница 95: ...Sequences This chapter explains sequences ...
Страница 153: ...Maintenance This chapter describes how to clean and inspect the PBZ ...
Страница 156: ...156 PBZ Maintenance ...
Страница 157: ...Appendix A List of Factory Default Settings B Troubleshooting C List of Error Messages D Options ...