2-6
PAT-T 4kW
2.5
Connecting the Power Cable
This product is designed as an equipment of IEC Overvoltage Category II (energy-
consuming equipment supplied from the fixed installation).
This product does not come with a power cable. In case the optional power cable
(AC8-4P4M-M6C) for three-phase input is used, attach crimping terminals that
comply with the terminal screws on the switchboard and connect the power cable
firmly so that it does not come loose. The crimping terminals connecting to the PAT
are already assembled.
Fig.2-3
Optional po
w
er cable
Exercise the following precautions when preparing the power cable.
Possible electric shock.
•
Turn off the switchboard breaker (switch that cuts off the power
supply from the switchboard) before making the connection.
•
This product is an IEC Safety Class I equipment (equipment with a
protective conductor terminal). Be sure to ground the product to
prevent electric shock.
•
Connect the
ground terminal to earth ground.
•
Be sure to have a qualified engineer connect the power cable to the
switchboard.
•
Inside the po
w
er s
u
pply, an appropriate protective circ
u
it is connected to
the inp
u
t terminal. Be s
u
re to connect the
w
ires correctly by follo
w
ing the
instr
u
ction for the connection of inp
u
t phase mode either in single-phase
inp
u
t mode or three-phase inp
u
t mode.
•
The POWER switch of the product can be used to disconnect the product from
the AC line in an emergency. Provide adequate space around the POWER switch
so that the POWER switch can be turned off at any time.
See
Three-phase inp
u
t po
w
er cable
(AC
8
-4P4M-M6C)
WARNING
CAUTION
Содержание PAT160-25T
Страница 17: ...General Description This chapter describes an overview and describes the features...
Страница 22: ...1 6 PAT T 4kW...
Страница 34: ...2 12 PAT T 4kW...
Страница 73: ...External Control This chapter describes external analog control and remote monitoring using the J1 connector...
Страница 106: ...6 12 PAT T 4kW...
Страница 158: ...7 52 PAT T 4KW...
Страница 159: ...Maintenance This chapter describes maintenance such as cleaning inspecting and calibrating...
Страница 171: ...Specifications This chapter describes the specifications and gives the dimensions of the PAT...