CURR
Sets the current.
Command
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]
{<numeric>|MINimum|MAXimum}
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]? [{MINimum|MAXimum}]
Parameter
Value 0 % to 105 % of the rated output current
(The default value is 105 % of the rated output current.)
An SCPI error (-221, "Settings conflict") occurs if
is not set to
NONE.
Unit
A
For the setting that is applied when *RST is sent, see
.
Response
Returns the current setting in the NR3 form in response to CURR?. If the current value is set
using external input, the specified current is returned.