INIT:CONT:SEQ3
INIT:CONT:NAME ACQ
Sets the measurement conditions of the sequence 3 group in sequence operation auto
continue mode.
When the sequence operation auto continue mode is turned on
If the trigger source parameter is set to IMM, the measurement starts immediately. When
the operation is complete, a new measurement automatically starts.
If the parameter is set to BUS, the measurement starts after receiving a software trigger.
When the operation is complete, the PAT-T waits for the next trigger.
When the sequence operation auto continue mode is turned off
The measurement currently in progress continues unless ABOR is sent. New
measurements are not automatically continued.
Command
INITiate:CONTinuous:SEQuence3 {ON|OFF|1|0}
INITiate:CONTinuous:SEQuence3?
INITiate:CONTinuous:NAME ACQuire,{ON|OFF|1|0}
INITiate:CONTinuous:NAME?
Parameter
Value ON(1) Auto continue mode on
OFF(0) Auto continue mode off (default)
For the setting that is applied when *RST is sent, see
.
Response
Returns the sequence operation auto continue mode setting in the NR1 form in response to
INIT:CONT:SEQ3? / INIT:CONT:NAME ACQ.