PAD-L
Setup
1-9
WARNING
•
For connections involving the input terminals of the main power
transformer, only change those of the AWG10 wiring. Do not
modify the AWG26 wiring, CR snubber, or varistor mounting.
Modifying the AWG26 wiring connection will destroy internal
circuits such as the sub-transformer.
4.
Replace the input fuse with a fuse meeting the input power voltage.
The input fuse must be rated to handle the input power voltage.
Table 1-2
IIIIn
n
n
npp
p
puu
u
utttt ffffuu
u
usss
see
e
e
PAD
16-100L
PAD
35-50L
PAD
35-60L
PAD
60-35L
PAD
110-20L
PAD
250-8L
For 200 V
30 A
For 100 V
60 A
50 A
50 A
5.
Turn over the input voltage mark-plate on the rear panel.
GND
AC 200V
50/60Hz
Input voltage mark-plate
AC input terminal board
Mounting screw M2.6 x2
INPUT
GND
AC 100V
50/60Hz
INPUT
Fig. 1-5
Input voltage mark-plate
6.
Replace the top panel.
Содержание PAD-L III Series
Страница 4: ...II PAD L...
Страница 16: ...P 6 Preface PAD L...
Страница 17: ...1 1 PAD L Setup 1 1 Chapter 1 Setup Describes the necessary procedure from unpacking to preparation before use...
Страница 40: ...2 12 Precautions and Preparations for Use PAD L...
Страница 86: ...4 34 Applied Operation PAD L...
Страница 93: ...PAD L Names and Functions of Controls 5 7 22 25 23 26 24 21 Fig 5 6 PAD L series PAD250 15L rear panel...
Страница 95: ...PAD L Names and Functions of Controls 5 9 22 26 23 25 21 24 28 29 Fig 5 8 PAD L series PAD110 60L rear panel...
Страница 98: ...5 12 Names and Functions of Controls PAD L...
Страница 130: ...7 14 Specifications PAD L...