3
KHA3000
65
Connecting External Signals
W
iring M
easurem
ent Circuits
EXT CLAMP (only for options)
Connect optional current sensors. This connection terminal is provided at each of the L1, L2,
and L3 input sections.
These terminals are enabled when "Sensor" is selected in the "Curr Input Terminal" in the
setting menu. The type and serial number of the current sensor will be displayed in the option
data in the system setting screen.
* At present, no optional current sensors are available although they are to be released in the
future.
Automatic recognition of optional current sensors
This product automatically recognizes any optional current sensors connected to the EXT
CLAMP terminals when the power is turned on. If a current sensor is replaced or newly
connected with the POWER switch turned on, the current sensor cannot be automatically
recognized.
- Whenever a current sensor is replaced or newly connected with the POWER switch turned
on, the user should take an action to make this product recognize the sensor.
- The operation to make this product recognize the current sensor (i.e., to press the "Update
Opt. Information" key) is performed in system setup.
- If any current sensor connected to this product has not been recognized, a dialog message
to prompt the operation to make the system recognize the current sensor (i.e., pressing the
"Update Opt. Information" key) will be presented when the "Curr Input Terminal" menu is
selected or when the test is started.
- If the power is turned on again after the current sensor is connected, the current sensors
connected will be automatically recognized.
EXT SYNC INPUT
EXT SYNC INPUT is an external synchronizing signal input terminal. The maximum input
voltage is 5 Vpeak. EXT SYNC INPUT is enabled when EXT is selected for
“
PLL Source
”
in
the setting menu.
See
See
See
+5 V
EXT SYNC INPUT
10 k
Ω
Содержание KHA3000
Страница 20: ...20 KHA3000...
Страница 35: ...Overview This chapter provides an overview of this product and explains the options for the product 1...
Страница 42: ...42 KHA3000 Options...
Страница 49: ...Wiring Measurement Circuits This chapter explains the wiring of measurement circuits and the control circuit 3...
Страница 70: ...70 KHA3000 Connecting Communication Cables...
Страница 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Страница 91: ...Setting the Harmonic Test This chapter explains how to set the test conditions for the harmonic current test 5...
Страница 120: ...120 KHA3000 Controlling the AC Power Supply...
Страница 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Страница 247: ...Other Measurements This chapter explains basic measurement FFT analyzer and in rush current measurement 11...
Страница 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Страница 352: ...352 KHA3000 Cleaning and Checking...
Страница 353: ...Specifications This chapter explains the specifications of this product 15...
Страница 390: ...390 KHA3000...
Страница 391: ......