Measurement Procedure

Set  the  necessary  measurement  conditions  in  advance  by  referring  to  the

TOS3200 user’s manual.
The connection between the EUT and the probe/test lead varies depending on

the type of touch current to be measured and the EUT class. See Table 4-1, “Test

lead connections for the type of TC measurement”,  chapter 4,  “TC and PCC

Tests” in the TOS3200 user’s manual. When referring the user’s manual, read

“test lead A” as “cable A.”
Connect test lead B to the test point on the EUT. For details, see “Connecting the

Test lead (TL21-TOS)”, chapter 4, “TC and PCC Tests” in the TOS3200 user’s

manual.

Fig.5

Connection example for touch current measurement

Starting the measurement

There  is  a  danger  of  electric  shock.  Never  touch  the
contact pin while the DANGER LED is illuminated.

When  you  connect  the  probe  to  the  TOS3200,  the  READY  indicator  on  the

display turns off. Even though, you can start the test using the START switch on

the probe.
Keep  holding  down  the  START  switch.  If  you  release  the  switch,  the

measurement  is  aborted.  The  DANGER  LED  illuminates  while  the

measurement  is  in  progress.  The  DANGER  LED  illuminates  (or  turns  off)

simultaneously with the DANGER lamp on the TOS3200.
When you connect the probe to the TOS3200, the START switch on the panel is

disabled. The STOP switch remains enabled.

EUT

To the test point

To the test point

WARNING