Component Names and Functions

No. Name

Description

1

Red Area

This part holds the contact pin to which high volt-
age is output.

2

Contact Pin

High voltage is delivered to this pin. The pin 
emerges from the red area when you depress the 
slide lever. The test probe is used with this pin 
contacting a test point of the DUT.

3

Start Switch 1 (trigger)

This is used to begin testing. Depress the trigger 
while grasping the test probe grip with the fingers 
of your dominant hand.

4

Start Switch 2

This is used to begin testing. Press the switch 
with a finger from your non-dominant hand. This 
hand should be free, since the other hand holds 
the test probe.

5

H.V ON LED

This red LED indicates that high voltage is being 
applied to the contact pin. It turns on or off in 
response to the on/off status of the DANGER 
(TEST ON or H.V ON) lamp of the Tester.

6

Grip

Grasp this part to use the test probe. Grasp with 
your dominant hand.

7

Slide Lever

Depressing this lever causes the contact pin to 
emerge from the red area. Be sure to depress it with 
the fingers of the hand grasping the test probe grip.

8

Remote Control Cable

Connect to the REMOTE terminal.

9

High Voltage Cable (red)

Connect to the high voltage terminal.

10

LOW Test Leadwire (black)

Connect to the LOW (GND) terminal

5

6

7

1

2

3

4

9

8

10

Содержание HP01A-TOS

Страница 1: ...electric shock that may result in serious injury or death Never perform testing when leaning on your elbows at the test bench or in other areas Do not fasten the slide lever of the test probe such th...

Страница 2: ...cting electric shock Modification may cause electric shock Never fasten the slide lever to the grip such that the contact pin is exposed Do not lock start switch 1 and or start switch 2 to the on stat...

Страница 3: ...ep this manual near the product to allow for easy reference by the operator when necessary About Manuals Intended readers These manuals are intended for users of this product and their instructors The...

Страница 4: ...sor in Charge of Operations 3 About Manuals 3 Inspection Before Starting 5 Stop Use in the Event of Failures 5 Component Names and Functions 6 Connecting to the Tester 7 Starting and Stopping Tests 8...

Страница 5: ...V ON lamp on the Tester The H V ON LED remains lit even when start switch 1 or 2 is turned off The H V ON LED remains lit even when the contact pin is retracted into the red area Item Description of I...

Страница 6: ...witch with a finger from your non dominant hand This hand should be free since the other hand holds the test probe 5 H V ON LED This red LED indicates that high voltage is being applied to the contact...

Страница 7: ...r 4 Connect the high voltage cable to the high voltage terminal of the Tester Cable Lock The cable lock guard is not used Insert the tester side end of the test leadwire fixture into the terminal and...

Страница 8: ...g test Disabled Press start switches 1 and2 Stopping test Press the STOP switch Release your finger from start switch 1 or 2 CAUTION Initiation of contact or removal from the DUT while high voltages a...

Страница 9: ...t lit g 11Release your fingers from the slide lever to retract the contact pin into the red area h 12Release the finger of your non dominant hand from start switch 2 i 13Remove the test probe from the...

Страница 10: ...ntly wipe the probe Periodic Inspection The remote control cable high voltage cable and LOW test leadwire are consum ables We recommend a inspection by one of our authorized service engineers once a y...

Страница 11: ...d to the contact pin during testing It also includes areas where high volt age is electrostatically or electromagnetically induced by application of high voltages 2 Rated maximum voltage The rated max...

Страница 12: ...ery effort has been made to ensure the accuracy of this manual However if you have any questions or find any errors or omissions please contact your Kikusui agent or distributor After you have finishe...

Отзывы: