Specifications | General specifications
General specifications
Item
BIM1030
BIM1100
Environment
Installation location
Indoors, 2000 m or less
Spec guaranteed
range
Temperature
18 °C to 28 °C (-4 °F to 158 °F)
Humidity
20 %rh to 80 %rh (no condensation)
Operating range
Temperature
0 °C to 40 °C (32 °F to +122 °F)
Humidity
20 %rh to 80 %rh (no condensation)
Storage range
Temperature
-10 °C to 60 °C (-4 °F to 158 °F)
Humidity
90 %rh or less (no condensation)
Power supply
Input voltage range
85 Vac to 264 Vac (100 Vac to 240 Vac)
Input frequency range
47 Hz to 63 Hz
Rated power
30 VA
Isolation voltage
±300 V max
±1000 V max
Insulation resistance (between AC LINE and chassis)
30 MΩ or more (500 Vdc)
Withstanding
voltage
Between the AC LINE and the chassis 1500 Vac for 1 minute, 10 mA or less
Between all the measurement termi-
nals and the chassis
2000 Vdc for 1 minute, 1 mA or less
Between all the measurement termi-
nals and SIGNAL I/O
2000 Vdc for 1 minute, 1 mA or less
External dimensions
Weight
Approx. 3 kg (6.6 lbs)
Accessories
Electromagnetic compatibility (EMC)
*1
*2
*1.
Does not apply to specially ordered or modified products.
*2.
Limited to products that have a CE mark.
Complies with the requirements of the following directive and
standards.
EMC Directive 2014/30/EU
EN 61326-1 (Class A
*3
)
EN 55011 (Class A*3, Group 1
*4
)
EN 61000-3-2
EN 61000-3-3
Applicable under the following conditions
The maximum length of all cabling and wiring connected to
the product must be less than 2.5 m.
Shielded cables are used to connect to SIGNAL I/O.
TL01-BIM clip-type four-wire test leads are used.
Electrical discharges are applied only to the DUT.
*3.
This is a Class A instrument. This product is intended for use in an industrial environment. This product may cause
interference if used in residential areas. Such use must be avoided unless the user takes special measures to reduce
electromagnetic emissions to prevent interference to the reception of radio and television broadcasts.
*4.
This is a Group 1 instrument. This product does not generate and/or use intentionally radio-frequency energy, in the form of
electromagnetic radiation, inductive and/or capacitive coupling, for the treatment of material or inspection/analysis purpose.
Safety*1
Complies with the requirements of the following directive and
standards.
Low Voltage Directive 2014/35/EU*2
EN 61010-1 (Class I
*5
, Pollution Degree 2
*6
)
EN 61010-2-030
EN 61010-031
*5.
This is a Class I instrument. Be sure to ground this product's protective conductor terminal. The safety of this product
is guaranteed only when the product is properly grounded.
*6.
Pollution is addition of foreign matter (solid, liquid or gaseous) that may produce a reduction of dielectric strength or
surface resistivity. Pollution Degree 2 assumes that only non-conductive pollution will occur except for an occasional
temporary conductivity caused by condensation.