Getting Started
1
Keysight X8712A IoT Device Battery Life Optimization Solution User’s Guide
31
Settings
Description
Zones
Automatically divides the Avg
Channel 1 trace into equal
‘zones’ determined by the
number of zones.
Show Zones
Divides the main chart into a number of regions selected in the
drop-down list, from 1 to 20. Default: 10
Sample & Duration
Configures the parameters
associated with data
acquisition.
Sample Size
Define the number of points in a measurement where the minimum
point is 1. The range is dependent on the number of channels below:
– One channel: up to 512K points.
– Two channels: up to 256K points.
– Three channels: up to 128K points.
– Four channels: up to 64K points
where K equals 1024.
Time Interval (s)
Define the time period between samples, where the maximum is 0.1
seconds. The minimum time interval is dependent on the number of
channels below:
– One channel: 5.12 microseconds.
– Two channels: 10.24 microseconds
– Three or four channels: 20.48 microseconds
Time Span (s)
The product of the Sample Size and Time Interval.