Crosstalk Measurement
In audio systems with more than one channel, it is common for a signal in one channel to
appear at the output of another channel at a reduced level. Crosstalk refers to this signal
leakage across channels, and it is expressed in dB (default) or as a percentage.
Crosstalk is a measurement of the ratio of the signal amplitude in an unused channel
relative to that of a channel driven with a signal. The unused channels should be grounded
or set to an appropriate bias point. Crosstalk is largely due to capacitive coupling between
the channel conductors in the device and generally varies with frequency. Crosstalk can be
computed as follows.
Phase measurement function mode is applicable for the analog, digital, and
Bluetooth
audio analyzer. The settings for crosstalk measurement are displayed as shown in
, and
for the analog, digital, and
Bluetooth
analyzer
respectively.
Figure 4-48
Analog Analyzer > Functions > X-Talk menu page 1
Crosstalk
20 Log10
rms value of signal measured
rms value of signal driven
---------------------------------------------------------
⎝
⎠
⎛
⎞
=
Содержание U8903B
Страница 1: ...Keysight U8903B Performance Audio Analyzer User s Guide ...
Страница 16: ...16 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 32: ...32 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 42: ...1 Getting Started 42 Keysight U8903B User s Guide LCD display Figure 1 4 U8903B LCD display 2 3 1 4 ...
Страница 96: ...2 Operation and Features 96 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 278: ...5 Graph Analysis 278 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 290: ...6 Sweep Function 290 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 406: ...7 Test Sequence Application 406 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 418: ...9 Characteristics and Specifications 418 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 574: ...A Appendixes 574 Keysight U8903B User s Guide Figure A 11 Bluetooth generator sine waveform generation ...