Recommended test equipment
The test equipment recommended for the calibration and adjustment is listed
below. If the exact instrument is not available, substitute the calibration standards
of equivalent requirement(s).
Table 5-2
List of equipment
Equipment
Requirement(s)
Recommended model
Purpose
Used
Digital Multimeter
(DMM)
8½ digits resolution
Keysight 3458A
Measures accurate DC
voltage
PV
[a]
C
[b]
Oscilloscope
– Sensitivity: 1 mV
– Bandwidth limit: 20 MHz
– Probe:
– 1:1 with RF tip
– 10:1 with RF tip for > 50 V
Keysight DSO8064A or
Infiniium equivalent
Displays transient
response. Displays ripple
and noise waveform.
PV
Differential Amplifier
– Bandwidth > 20 MHz (bandwidth
limit : 20 MHz)
– AC coupling
– Amplifier x10
LeCroy DA1855A
Signal conditioning
preamplifiers
PV
RMS Volt Meter
20 Hz to 20 MHz
R&S URE3
Measures RMS noise
PV
50
Ω
Feed-through
Termination
Pomona
Loads termination to
match output impedance
of the differential amplifier
PV
AC/DC Current
Converter
Tektronix TCP305 &
Tektronix TCPA300
Converts current to
voltage
Electronic Load
– Voltage range: 30 Vdc
– Current range: 5 A
– Open and Short switches
– Transient On/Off
Keysight 6060B
Measures load and line
regulations and transient
response time
PV
AC Power Source
Capable of supplying
90 Vac to 250 Vac
Keysight 6813B
Functions as variable
voltage transformer
PV
Содержание U8001A
Страница 8: ...8 Keysight U8001A U8002A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 12: ...12 Keysight U8001A U8002A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 14: ...14 Keysight U8001A U8002A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 69: ...Service Guide 5 Keysight U8001A U8002A User s and Service Guide 69 Figure 5 4 Graph of load transient response ...