General Information
2
Keysight U8001A/U8002A User’s and Service Guide
29
Output Connections
Voltage drops
The load wires must be large enough to avoid excessive voltage drops due to the
impedance of the wires. In general, if the wires are heavy enough to carry the
maximum short circuit current without overheating, excessive voltage drops will
not be a problem. The voltage drops across the load wires should be limited to
less than 2 V. Refer to the American Wire Gauge (AWG) standard to calculate the
voltage drop for some commonly used AWG wire copper.
WARNING
Before attempting to connect wires to the front output terminals, make sure
to turn off the power supply first to avoid damage to the circuits being
connected.