1
Calibration Procedures
94
Keysight U3606B Service Guide
8
Verify the constant voltage adjustments using the
Table 1-33
Constant voltage output adjustment
Adjustment item
Output voltage
OUt-L
AUTO calibration
OUt-H
AUTO calibration
LOAd1
[a]
[a] See
for information on how to connect the LOAd adjustment item.
30 V with additional 30
Ω
(30 W) fixed resistor
LOAd2
8 V with additional 2.67
Ω
(24 W) fixed resistor
LOAd3
[a][b]
[b] S2 range with a lighter load.
8 V with additional 300
Ω
(3 W) fixed resistor
LOAd4
1000 mV with additional 0.33
Ω
(3 W) fixed resistor
Fixed resistor