1
Calibration Procedures
88
Keysight U3606B Service Guide
7
Press
to start the adjustment. The word CALib starts flashing to indicate
that the calibration is in progress.
– Successful completion of the adjustment is indicated by a short beep and
the display briefly showing PASS.
– An adjustment failure is indicated by a long beep, the display showing FAiL
along with a calibration error number. Correct the problem and repeat this
procedure.
8
Repeat
through
for each gain adjustment point shown in
.
9
Verify the capacitance gain adjustments using the
Table 1-30
Capacitance gain adjustment
Adjustment item
Input capacitance
OPEn
Input terminals open (remove any test leads or shorting
plugs from the input terminals)
0.400 nF
0.4 nF
1.000 nF
1 nF
10.00 nF
10 nF
100.0 nF
100 nF
1.000
μ
F
1
μ
F
10.00
μ
F
10
μ
F
100.0
μ
F
100
μ
F
1000
μ
F
1000
μ
F
10000
μ
F
10000
μ
F
OUT
SBY