Calibration Procedures
1
Keysight U3606B Service Guide
85
6
Enter the actual applied input.
7
Press
to start the adjustment. The word CALib starts flashing to indicate
that the calibration is in progress.
– Successful completion of the adjustment is indicated by a short beep and
the display briefly showing PASS.
– An adjustment failure is indicated by a long beep, the display showing FAiL
along with a calibration error number. Correct the problem and repeat this
procedure.
8
Repeat
through
for each gain adjustment point shown in
.
9
Verify the DC current gain adjustments using the
NOTE
Always complete tests in the same order as shown in
.
Table 1-28
DC current gain adjustment
Adjustment item
Input current
OPEn
Input terminals open (remove any test leads or shorting
plugs from the input terminals)
10.0000 mA
10 mA
100.000 mA
100 mA
1000.00 mA
1000 mA
OUT
SBY