1
Calibration Procedures
70
Keysight U3606B Service Guide
Figure 1-18
SECUR pads location
4
Apply power and turn on the instrument.
5
The display will show the message “CALib”. The instrument is now unsecured.
(The security code resets to the factory default value.)
6
Turn off the instrument and remove the power cord.
7
Remove the temporary short installed in
.
8
Reassemble the instrument.