Calibration Procedures
1
Keysight U3606B Service Guide
47
Frequency voltage gain verification test
1
Connect the calibrator to the front panel
V
(red) and
LO
(black) input terminals
(see
Figure 1-7, “Test setup for frequency voltage gain verification,”
2
Press
[Hz ms %]
to select the frequency function.
3
Select each range in the order shown below. Provide the indicated input
frequency and voltage. Compare the measurement results and the appropriate
test limits in
(be certain to allow for appropriate source settling when
using the Fluke 5520A).
Frequency current gain verification test
1
Connect the calibrator to the front panel
I
(red) and
LO
(black) input terminals
(see
Figure 1-8, “Test setup for frequency current gain verification,”
2
Press
[Hz ms %]
to select the frequency function.
Table 1-8
4-wire resistance gain verification test
Input resistance
Range
Error from nominal 1-year
1
Ω
[a]
[a] Calibrator ohms zero cal. every 12 hours ± 1
o
C
1
Ω
±0.0028
Ω
10
Ω
[a]
10
Ω
±0.012
Ω
100
Ω
100
Ω
±0.12
Ω
1000
Ω
1000
Ω
±1.2
Ω
Table 1-9
Frequency voltage gain verification test
Input frequency
Input voltage
Range
Error from nominal 1-year
1 kHz
1 V
1.999 kHz
±0.00026 kHz
Содержание U3606B
Страница 1: ...Keysight U3606B Multimeter DC Power Supply Service Guide ...
Страница 8: ...8 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 12: ...12 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 14: ...14 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 104: ...1 Calibration Procedures 104 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...