8
Characteristics and Specifications
DC Power Supply Specifications
258
U3606A User’s and Service Guide
Square-wave output specifications
Remote sensing capability:
Output can maintain specifications for
•
Range S1 (30 V/1 A): Up to a 0.75-volt drop per load lead.
•
Range S2 (8 V/3 A): Up to a 0.5-volt drop per load lead.
Voltage programming speed:
(excludes command
processing time)
Full load
•
Up: 300 ms
•
Down: 400 ms
No load
•
Up: 300 ms
•
Down: 400 ms
Table 8-9
Square-wave output specifications
Parameter
Specifications
Amplitude accuracy
± (offset)
Range S1
0.2 V
Range S2
0.2 V
Amplitude resolution
Range S1
1 mV
Range S2
1 mV
Frequency accuracy
± (% of frequency s offset)
Range
(27 steps
[1]
)
0.005% + 1 count
Frequency resolution
0.01 Hz
Duty cycle accuracy
± (% of duty cycle setting)
Range
(256 steps :
0.39% to 99.60%)
0.4%
[2][3]
Duty cycle resolution
0.39%
[3]
[3][4]
± (offset)
Range
(256 steps:
1/frequency)
Duty cycle/frequency
Pulse width resolution
Range/256
[1] Available frequencies: 0.5, 2, 5, 6, 10, 15, 25, 30, 40, 50, 60, 75, 80, 100, 120, 150, 200, 240, 300, 400, 480, 600, 800, 1200, 1600,
2400, 4800 (Hz)
Содержание U3606A
Страница 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Страница 2: ......
Страница 12: ...12 U3606A User s and Service Guide ...
Страница 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Страница 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Страница 198: ...5 Verification and Performance Tests Additional Verification Tests 176 U3606A User s and Service Guide ...
Страница 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Страница 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...