5
Verification and Performance Tests
Additional Verification Tests
172
U3606A User’s and Service Guide
Additional Verification Tests
Optional capacitance gain verification test
1
Connect the calibrator to the front panel
(red) and
LO
(black) input
).
2
Press
to select the capacitance function.
3
Select each range in the order shown below. Provide the indicated input
capacitance. Compare the measurement results to the appropriate test
limits shown in
. (Be certain to allow for appropriate source
settling when using the Fluke 5520A.)
Table 5-15
Optional capacitance gain verification test
Input capacitance
Range
Error from nominal 1 year
0.4 nF
1 nF
±0.016 nF
1 nF
10 nF
±0.06 nF
10 nF
100 nF
±0.6 nF
0.1
μ
F
1
μ
F
±0.006
μ
F
1
μ
F
10
μ
F
±0.06
μ
F
10
μ
F
100
μ
F
±0.6
μ
F
100
μ
F
1000
μ
F
±6
μ
F
1000
μ
F
10000
μ
F
±70
μ
F
1 nF
1 nF
±0.028 nF
10 nF
10 nF
±0.15 nF
0.1 nF
100 nF
±0.501 nF
1
μ
F
1
μ
F
±0.015
μ
F
10
μ
F
10
μ
F
±0.15
μ
F
100
μ
F
100
μ
F
±1.5
μ
F
Содержание U3606A
Страница 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Страница 2: ......
Страница 12: ...12 U3606A User s and Service Guide ...
Страница 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Страница 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Страница 198: ...5 Verification and Performance Tests Additional Verification Tests 176 U3606A User s and Service Guide ...
Страница 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Страница 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...