Utility Menu Summary
120
U3606A User’s and Service Guide
Performing a self-test
A power- on self- test occurs immediately when you turn on the
instrument. This limited test assures you that the U3606A is operational.
The following procedure shows you how to perform a more extensive
self- test. A complete self- test performs a series of internal tests, and may
take up to 30 seconds to complete. For remote interface operation, refer
to the
*TST
command in the
U3606A Programmer’s Reference
.
1
To access the utility menu, press
Shift
>
Utility
.
2
The first utility menu item (Error) will be shown in the upper
secondary display.
3
Press
or
until the menu item “SELF.t” is shown in the upper
secondary display.
4
Press
or
to select “YES”.
5
Press
Shift
>
Save
to save. The instrument will automatically exit the
utility menu and execute the self- test.
6
If the self- test is successful, the U3606A will return to normal
operation.
C A U T I O N
Before performing the complete self-test, ensure that all test leads are
removed from the input and output terminals.
Содержание U3606A
Страница 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Страница 2: ......
Страница 12: ...12 U3606A User s and Service Guide ...
Страница 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Страница 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Страница 198: ...5 Verification and Performance Tests Additional Verification Tests 176 U3606A User s and Service Guide ...
Страница 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Страница 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...