5
Disassembly and Repair
88
Keysight U3401A User’s and Service Guide
To replace a current input fuse
Both the mA and the A current input terminals are fuse protected. The fuse for the
mA input terminal is located on the front panel (see
). The fuse is a 0.63 A,
500 V fuse (refer to
). If you determine that the fuse is faulty, replace it
with one of the same size and rating.
The fuse for the A current input terminal is located inside the multimeter (see
) and requires partial disasembly of the multimeter. The fuse is a 25 A,
440 V fast-acting fuse (refer to
). If you determine that the fuse is faulty,
replace it with one of the same size and rating.
Electrostatic Discharge (ESD) precautions
Almost all electrical components can be damaged by electrostatic discharge
(ESD) during handling. Component damage can occur at electrostatic discharge
voltages as low as 50 volts.
The following guidelines will help prevent ESD damage when servicing the
instrument or any electronic device.
– Disassemble instruments only in a static-free work area.
– Use a conductive work area to reduce static charges.
– Use a conductive wrist strap to reduce static charge accumulation.
– Minimize handling.
– Keep replacement parts in original static-free packaging.
– Remove all plastics, foam, vinyl, paper, and other static-generating materials
from the immediate work area.
– Use only anti-static solder suckers.
Table 5-2
Type of current input fuse
Type of fuse
Part number
0.63 A, 500 V fuse
2110-1432
25 A, 440 V fast-acting fuse
2110-1431
Содержание U3401A
Страница 1: ...Keysight U3401A 4 1 2 Digit Dual Display Multimeter User s and Service Guide ...
Страница 12: ...12 Keysight U3401A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 16: ...16 Keysight U3401A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 70: ...3 Application Tutorial 70 Keysight U3401A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 82: ...4 Performance Test 82 Keysight U3401A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...