Performance Test
4
Keysight U3401A User’s and Service Guide
81
AC Current Verification Test
Configuration: AC current
1
Select the AC current function.
2
Select each range in the order shown in
. Provide the input current
and frequency indicated. Compare measurement results to the appropriate test
. (Be certain to allow for appropriate source settling).
Table 4-8
AC current verification test
Function
Current
Input frequency
Range
Error from nominal one year
AC current
50.00
μ
A
1 kHz
500
μ
A
±0.45
μ
A
450.00
μ
A
1 kHz
500
μ
A
±2.45
μ
A
0.5000 mA
1 kHz
5 mA
±4.5
μ
A
4.5000 mA
1 kHz
5 mA
±24.5
μ
A
5.000 mA
1 kHz
50 mA
±45
μ
A
45.000 mA
1 kHz
50 mA
±0.245 mA
50.00 mA
1 kHz
500 mA
±0.45 mA
450.00 mA
1 kHz
500 mA
±2.45 mA
0.5000 A
1 kHz
5 A
±5.5 mA
1.000 A
1 kHz
10 A
±35 mA
4.5000 A
1 kHz
5 A
±25.5 mA
9.000 A
1 kHz
10 A
±75 mA