U3042AE16
N5222/42A/B RF Interface Cable Connections (U3021-60077)
N5222/42A/B RF Interface Cable Connections (U3021-60077)
illustrates the cable configuration of the test set to the analyzer.
The cables have been supplied the kit (U3021-60077).
Cable Connection Procedure
1. Remove the SOURCE OUT to CPLR THRU and RCVR IN to CPLR ARM jumpers (x8) on the
analyzer. The RCVR R1 - R4 to SOURCE OUT reference loop jumpers (x4) remain on the front
panel.
2. Connect the RF interconnect cables from the PNA or PNA-X to the test set in the order listed.
As you are connecting each cable, torque to 8 in-lb. Refer to
and
Over-torque will cause damage to the test set and may cause connectors
to spin or become loose.
Refer to
“Interconnect Cable Verification” on page 2
.
Table 9
PNA or PNA-X RF Interface Cable Connection (U3021-60077)
Numeric Order
RF Cables
From: PNA-X
To: Test Set
1
U3042-20122
Port 1 CPLR THRU
Port 5 CPLR THRU
2
U3042-20120
Port 1 SOURCE OUT
Port 5 SOURCE IN
3
U3042-20119
Port 1 CPLR ARM
Port 5 CPLR ARM
4
U3042-20121
Port 1 RCVR A IN
Port 5 RCVR OUT
5
U3042-20123
Port 3 CPLR ARM
Port 6 CPLR ARM
6
U3042-20123
Port 3 RCVR C IN
Port 6 RCVR OUT
7
U3042-20124
Port 3 SOURCE OUT
Port 6 SOURCE IN
8
U3042-20124
Port 3 CPLR THRU
Port 6 CPLR THRU
9
U3042-20123
Port 4 RCVR D IN
Port 7 RCVR OUT
10
U3042-20123
Port 4 CPLR ARM
Port 7 CPLR ARM
11
U3042-20124
Port 4 SOURCE OUT
Port 7 SOURCE IN
12
U3042-20124
Port 4 CPLR THRU
Port 7 CPLR THRU
13
U3042-20126
Port 2 CPLR THRU
Port 8 CPLR THRU
14
U3042-20128
Port 2 SOURCE OUT
Port 8 SOURCE IN
15
U3042-20125
Port 2 RCVR B IN
Port 8 RCVR OUT
16
U3042-20127
Port 2 CPLR ARM
Port 8 CPLR ARM
22
User's and Service Guide U3042-90012