U2722A/U2723A Service Guide

2

Getting  Started

 

 

Introduction

Self-Test 

A brief power

-

on self

-

test occurs automatically whenever 

you turn on the instrument. This limited test assures that 
the instrument is capable of operation.

If the self

-

test fails, the power indicator on the front 

panel will start blinking. You can use the 

SYSTem:ERRor? 

query command from the remote interface to read the 
error messages. If repair is required, contact a Keysight 
Service Center. Please refer to the programmer’s reference 
guide for more information on the error messages.

Self

-

test may be initiated remotely by sending the 

*TST?

 

SCPI command to the instrument.

Keysight Measurement Manager

To start the self- test procedure using the Keysight 
Measurement Manager, perform the following steps.

1

Click 

Tools > Self-Test

.

C A U T I O N

Disconnect all connections from the output terminals before 
performing the self-test procedure. Ensure that the output terminals 
are not connected to any load before you turn on the U2722A/U2723A.

Содержание u2722a

Страница 1: ...Keysight U2722A U2723A USB Modular Source Measure Units Service Guide ...

Страница 2: ......

Страница 3: ......

Страница 4: ...n the separate agreement shall control Technology Licenses The hardware and or software described in this document are furnished under a license and may be used or copied only in accor dance with the terms of such license Restricted Rights Legend U S Government Restricted Rights Soft ware and technical data rights granted to the federal government include only those rights customarily provided to ...

Страница 5: ...sulation or reinforced insulation Alternating current AC Off supply Both direct and alternating current On supply Three phase alternating current Caution risk of electric shock Earth ground terminal Caution risk of danger refer to this manual for specific Warning or Caution information Protective conductor terminal Caution hot surface Frame or chassis terminal Out position of a bi stable push cont...

Страница 6: ... trained personnel who are aware of the haz ards involved should remove the instrument casing Always discon nect the power cable and any external circuits before removing the instrument casing Do not use the equipment if it does not operate properly Ensure that the equipment is inspected by qualified service personnel Do not install substitute parts or perform any unauthorized modifica tion to the...

Страница 7: ...ng temperature 0 C to 50 C Operating humidity 20 to 85 RH non condensing Storage temperature 20 C to 70 C Storage humidity 5 to 90 RH non condensing CAUTION The U2722A U2723A USB Modular Source Measure Units complies with the following safety and EMC requirements IEC 61326 1 2005 EN61326 1 2006 Canada ICES 001 2004 Australia New Zealand AS NZS CISPR11 2004 IEC 61010 1 2001 EN 61010 1 2001 2nd Edit...

Страница 8: ... with the Australia EMC Framework regulations under the terms of the Radio Communication Act of 1992 ICES NMB 001 indicates that this ISM device complies with the Cana dian ICES 001 Cet appareil ISM est confomre a la norme NMB 001 du Canada This instrument complies with the WEEE Directive 2002 96 EC mark ing requirement This affixed product label indicates that you must not dis card this electrica...

Страница 9: ...s electrical or electronic product in domestic household waste Product Category With reference to the equipment types in the WEEE directive Annex 1 this instrument is classified as a Monitoring and Control Instrument The affixed product label is shown as below Do not dispose in domestic household waste To return this unwanted instrument contact your nearest Keysight Technologies or visit www keysi...

Страница 10: ...er provides an overview of the service procedure for the U2722A U2723A which includes the service types offered by Keysight and instructions for obtaining services and sending the U2722A U2723A for service 3 Performance Verification and Adjustment This chapter describes the calibration procedures for the U2722A U2723A to ensure that the instrument is operating normally within the warranted specifi...

Страница 11: ...Services 14 Calibration Interval 14 Types of Service Available 14 Extended Service Contracts 14 Obtaining Repair Service Worldwide 15 Repackaging for Shipment 16 Cleaning 16 3 Performance Verification and Adjustment 17 Performance Verification and Adjustment 18 Recommended Test Equipment 18 Test Consideration 18 Performance Verification 19 Voltage Tests 19 Current Tests 21 Test Record Forms 24 Sou...

Страница 12: ...X U2722A U2723A Service Guide Contents Source Current Mode Programming and Readback Accuracy 25 Adjustment 26 ...

Страница 13: ...U2722A U2723A Service Guide XI List of Figures Figure 3 1 Voltage test setup no load condition 20 Figure 3 2 Current test setup no load condition 22 ...

Страница 14: ... 2 Test settings for source voltage mode programming and readback accuracy 19 Table 3 3 Test settings for source current mode programming and readback accuracy 21 Table 3 4 Source voltage mode programming and readback accuracy 24 Table 3 5 Source current mode programming and readback accuracy 25 ...

Страница 15: ... Modular Source Measure Units Service Guide 1 Specifications Product Characteristics 2 Product Specifications 4 This chapter specifies the characteristics environmental conditions and specifications of the U2722A U2723A ...

Страница 16: ... degree 2 For indoor use only STORAGE COMPLIANCE 20 C to 70 C Relative humidity at 5 to 90 RH non condensing SAFETY COMPLIANCE Certified with IEC 61010 1 2001 EN61010 1 2001 2nd Edition Canada CAN CSA C22 2 No 61010 1 04 USA ANSI UL 61010 1 2004 EMC COMPLIANCE IEC 61326 2002 EN61326 1997 A1 1998 A2 2001 A3 2003 Canada ICES 001 2004 Australia New Zealand AS NZS CISPR11 2004 SHOCK AND VIBRATION Test...

Страница 17: ...r the product s standard accessories unless otherwise specified Please take note that for the product the warranty does not cover Damage from contamination Normal wear and tear of mechanical components Manuals 1 Compatible with Microsoft Windows operating systems only 2 Requires a direct USB connection to the PC so the appropriate driver can be in stalled in the USB modular instrument or USB DAQ m...

Страница 18: ...hs at 25 C 3 C of output offset 1 μA 0 085 0 85 nA 100 pA 10 μA 0 085 8 5 nA 1 nA 100 μA 0 075 75 nA 10 nA 1 mA 0 075 750 nA 100 nA 10 mA 0 075 7 5 μA 1 μA 120 mA 0 1 100 μA 20 μA Voltage readback 12 months over USB with respect to the actual output at 25 C 3 C of output offset 2 V 0 075 1 5 mV 0 1 mV 20 V 0 05 10 mV 1 mV Current readback 12 months over USB with respect to the actual output at 25 ...

Страница 19: ... 5 nA 1 nA 100 μA 0 075 75 nA 10 nA 1 mA 0 075 750 nA 100 nA 10 mA 0 075 7 5 μA 1 μA 120 mA 0 1 100 μA 20 μA Voltage readback 12 months over USB with respect to the actual output at 23 C 5 C of output offset 2 V 0 075 1 5 mV 0 1 mV 20 V 0 05 10 mV 1 mV Current readback 12 months over USB with respect to the actual output at 23 C 5 C of output offset 1 μA 0 085 0 85 nA 100 pA 10 μA 0 085 8 5 nA 1 n...

Страница 20: ...t For resistive measurement 2 1 μA 170 0 10 μA 18 0 100 μA 6 0 1 mA 1 0 10 mA 1 0 120 mA 1 0 1 Drive 50 of 1 V or 10 V output with a resistive load Rise time is from 10 to 90 of program voltage change at maximum current Fall time is from 90 to 10 of program voltage change at maximum current 2 Measurements obtained are per default bandwidth setting Voltage Range Current range 1 μA 10 μA 100 μA 1 mA...

Страница 21: ...04 of range Current load regulation Current load regulation 0 04 100 μA Programming language Standard Commands for Programmable Instruments SCPI Recommended calibration interval One year 1 Measurements obtained are per default bandwidth setting NOTE All channels are isolated from the ground and from each other Isolation is 60 VDC Category I Three hours warm up time recommended The measurement accu...

Страница 22: ...8 U2722A U2723A Service Guide 1 Specifications THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...

Страница 23: ...14 Extended Service Contracts 14 Obtaining Repair Service Worldwide 15 Repackaging for Shipment 16 Cleaning 16 This chapter discusses the self test and self calibration procedure for the U2722A U2723A USB Modular Source Measure Units It also provides the information for returning your U2722A U2723A to Keysight for calibration or servicing If you have a defective module you can return it to Keysigh...

Страница 24: ...essages If repair is required contact a Keysight Service Center Please refer to the programmer s reference guide for more information on the error messages Self test may be initiated remotely by sending the TST SCPI command to the instrument Keysight Measurement Manager To start the self test procedure using the Keysight Measurement Manager perform the following steps 1 Click Tools Self Test CAUTI...

Страница 25: ...A Service Guide 11 2 Click Start on the Self Test form 3 Click OK to continue when the message box appears 4 Wait for a few minutes for the self test to complete 5 The result is displayed on the form once the self test has completed ...

Страница 26: ...e Guide 2 Getting Started Self Calibration To perform self calibration using the U2722A U2723A Keysight Measurement Manager perform the following steps 1 Click Tools Self Calibration 2 Click Start on the Self Calibration form ...

Страница 27: ...splayed on the form once the self calibration has completed CAUTION Disconnect all connections from the output terminals before performing the self calibration procedure Ensure that the output terminals are not connected to any load before you turn on the U2722A U2723A It is recommended that the U2722A U2723A is powered up for at least 45 minutes before performing self calibration ...

Страница 28: ...vals beyond two years for any application Keysight recommends that a complete readjustment should always be performed at the calibration interval This will ensure that the U2722A U2723A will remain within specifications for the next calibration interval The readjustment provides the best long term stability and accuracy Types of Service Available If your instrument fails during the warranty period...

Страница 29: ...ce or technical support information you can contact Keysight at one of the following telephone numbers In the United States 800 829 4444 In Europe 31 20 547 2111 In Japan 81 426 56 7832 You can also use our Web link for the information on contacting Keysight worldwide www keysight com find assist Or contact your Keysight representative Before shipping your instrument ensure that you acquire shippi...

Страница 30: ...riate packaging material for shipping Secure the container with strong tape or metal bands If the original shipping container is not available place your unit in a container with at least four inches of compressible packaging material around all sides for the instrument Use static free packaging materials to avoid additional damage to your unit Cleaning Clean the outer area of the instrument with ...

Страница 31: ...rformance Verification 19 Voltage Tests 19 Current Tests 21 Test Record Forms 24 Source Voltage Mode Programming and Readback Accuracy 24 Source Current Mode Programming and Readback Accuracy 25 Adjustment 26 The performance verification procedures described in this chapter verify that the U2722A U2723A USB Modular Source Measure Units is operating normally and is within its warranted specificatio...

Страница 32: ...ct equipment is not available substitute the performance verification standards of equivalent requirement s Test Consideration For optimum performance all procedures should comply with the following recommendations Ensure that the ambient temperature is stable and between 22 C and 28 C The ideal temperature for performing performance verification should be 25 C 1 C Make sure that the ambient relat...

Страница 33: ... and readback through the USB are within published specifications Test Settings 1 Use the highest resolution available on the DMM NOTE Test each output channel individually Refer to the appropriate test record form Table 3 2 Test settings for source voltage mode programming and readback accuracy Keysight 3458A settings1 Function Range DC voltage 1 V for U2722A U2723A drives 1 V 10 V for U2722A U27...

Страница 34: ...on the U2722A U2723A 2 Configure the selected output as shown in the following a Configure the NPLC to 10 b Configure the voltage range c Configure the current range d Program the clamp current Iclamp e Program the drive voltage Vdrive 3 Enable the output 4 Wait for a few seconds for the output to settle 5 Record the voltage reading on the DMM Vmeas 6 Read back the drive voltage over the USB 7 Rec...

Страница 35: ...he current programming and readback through the USB are within published specifications Test Settings 1 Use the highest resolution available on the DMM Table 3 3 Test settings for source current mode programming and readback accuracy Keysight 3458A settings1 Function Range DC current 1 μA for U2722A U2723A drives 1 μA 10 μA for U2722A U2723A drives 10 μA 100 μA for U2722A U2723A drives 100 μA 1 mA...

Страница 36: ...rn on the U2722A U2723A 2 Configure the selected output as shown in the following a Configure the NPLC to 10 b Configure the voltage range c Configure the current range d Program the clamp voltage Vclamp e Program the drive current Idrive 3 Enable the output 4 Wait for a few seconds for the output to settle 5 Record the current reading on the DMM Imeas 6 Read back the drive current over the USB 7 ...

Страница 37: ...n and Adjustment 3 U2722A U2723A Service Guide 23 9 Repeat step 2 through step 8 for each iteration to complete the performance verification for the selected output 10 Repeat step 1 through step 9 for the rest of the outputs ...

Страница 38: ... 4 Source voltage mode programming and readback accuracy Iteration Test point Voltage range Clamp current Current range Vmeas Vdrive Vrdbk Vmeas Limit 1 1 V 2 V 120 mA 120 mA 2 25 mV 2 10 mA 10 mA 3 1 mA 1 mA 4 100 μA 100 μA 5 10 μA 10 μA 6 1 μA 1 μA 7 10 V 20 V 120 mA 120 mA 15 mV 8 10 mA 10 mA 9 1 mA 1 mA 10 100 μA 100 μA 11 10 μA 10 μA 12 1 μA 1 μA ...

Страница 39: ...on Test point Current range Clamp voltage Voltage range Imeas Idrive Irdbk Imeas Limit 1 0 5 μA 1 μA 20 V 20 V 1 275 nA 2 2 V 2 V 1 275 nA 3 5 μA 10 μA 20 V 20 V 12 75 nA 4 2 V 2 V 12 75 nA 5 50 μA 100 μA 20 V 20 V 112 5 nA 6 2 V 2 V 112 5 nA 7 0 5 mA 1 mA 20 V 20 V 1 125 μA 8 2 V 2 V 1 125 μA 9 5 mA 10 mA 20 V 20 V 11 25 μA 10 2 V 2 V 11 25 μA 11 60 mA 120 mA 20 V 20 V 0 16 mA 12 2 V 2 V 0 16 mA ...

Страница 40: ... The following information describes the procedure for voltage and current adjustments for the U2722A U2723A using Keysight Measurement Manager revision 1 7 or above 1 Establish the connection between the U2722A U2723A and the PC Launch the Keysight Measurement Manager as shown below ...

Страница 41: ...ct Tools Source Measure Adjustment to start the adjustment 3 A message box will appear to inform that the adjustment operation will take approximately two hours During this period you will not be able to operate the U2722A U2723A Click Yes to continue with the adjustment operation ...

Страница 42: ...ibration During this process make sure that there is no connection on the instruments terminals If self test or self calibration fails cycle the U2722A U2723A power and repeat step 1 through step 5 In the unlikely event that self test or or self calibration repeatedly fails contact your nearest Keysight Sales Office 6 After the self test and self calibration have completed a message box shown in t...

Страница 43: ...voltage and current adjustments for channel 2 followed by voltage and current adjustments for channel 3 Follow the instructions on the message box to complete the adjustment operation 9 When the adjustment operation has completed a message box as shown in the following will appear Click OK to end the adjustment operation 10 A message box will appear as shown below prompting you to power cycle the ...

Страница 44: ...30 U2722A U2723A Service Guide 3 Performance Verification and Adjustment THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...

Страница 45: ...e EMC 2 safety 2 storage 2 connection cable See test connector output See output consideration test See test contact 10 14 15 contract service 14 15 current clamp 20 DC 21 drive 22 22 output ratings 4 programmable 18 programming 4 21 readback 4 21 22 range 20 22 source See source test 21 22 D DC current See current voltage 19 degree pollution See pollution degree Digital Multimeter 18 19 20 21 22 ...

Страница 46: ...ltage See current or voltage ratings output See output readback See current or voltage recommended test equipment 18 record form test See test Regulatory markings VI relative humidity See humidity relative remote interface 2 10 sense operating range See operating range remote sense repackaging for shipment 16 repair service 14 15 resolution 4 18 19 21 result test 18 rise fall time 6 S safety See c...

Страница 47: ...voltage 19 setup current 22 setup voltage 20 shock and vibration 2 voltage See voltage tools self calibration 12 self test 10 TST 10 types service See service U USB 4 19 20 22 Hi Speed 2 0 See Hi Speed USB 2 0 USBTMC 488 2 2 V vibration See shock and vibration voltage clamp 22 DC 19 drive 20 20 output ratings 4 output overshoot See output programmable 18 programming 4 19 21 range 20 22 readback 4 ...

Страница 48: ...34 U2722A U2723A Service Guide Index THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...

Страница 49: ...ope tel 31 20 547 2111 Japan tel 81 426 56 7832 fax 81 426 56 7840 Korea tel 080 769 0800 fax 080 769 0900 Latin America tel 305 269 7500 Taiwan tel 0800 047 866 fax 0800 286 331 Other Asia Pacific Countries tel 65 6375 8100 fax 65 6755 0042 Or visit Keysight World Wide Web at www keysight com find assist Product specifications and descriptions in this document are subject to change without notice...

Страница 50: ...This information is subject to change without notice Keysight Technologies 2009 2014 Edition 6 November 2014 U2722 90021 U2722 90021 www keysight com ...

Отзывы: