U1701B Quick Start Guide

Actions

Steps

To cycle through maximum, minimum, average, and 
present readings 

  • MAX

MIN

AVG

 or 

MAX AVG MIN

 annunciator 

will be turned on respectively to indicate which 
value is being displayed

Press 

momentarily

To exit the recording mode

Press 

 for

more than 1 s

To enable relative function

  •

Relative function shows the difference between 
the measured value and the offset reference value. 
The display may show a non-zero value due to the 
presence of test leads. Use the relative function to 
nullify the residual.

  •

Relative function can operate in both auto and 
manual ranging mode but the function cannot be 
set when an overload value exists.

  • REL

 annuciator will be displayed.

Press 

To renew the relative value

Press 

 again

To exit relative mode

Press 

 for

more than 1 s

To select manual range and to turn off the 

AUTO

 

annunciator

Press 

To step up a range at a time

Press 

 again

To select auto-range

  •

In auto range mode, the 

AUTO

 annunciator is 

displayed and the instrument will select an 
appropriate range for resolution if the reading is 
greater than the maximum available range. 

OL

 will 

be displayed.

  •

The instrument will select a lower range when the 
reading is less than 9% of full scale.

Press  

for

more than 1 s

To enable the tolerance mode and to set the display 
value as a standard reference

  • TOL

 annunciator will be displayed. 

  •

The tolerance will be displayed on the secondary 
display.

  •

The instrument range will be locked.

Press 

Содержание u1701b

Страница 1: ...Keysight U1701B Dual Display Handheld Capacitance Meter Quick Start Guide ...

Страница 2: ......

Страница 3: ... Calibration If anything is missing or damaged please contact the nearest Keysight Sales Office For more detailed information please refer to the Keysight U1701B Dual Display Handheld Capacitance Meter User s Guide on Keysight Web site www keysight com find handheld tools WARNING To avoid damage to the device do not exceed the input limit Do not apply voltage to input terminals Discharge the capac...

Страница 4: ...he meter and leads 3 Insert the capacitor legs into and input terminals respectively Ensure that the polarity of the capacitor s leg are correct 4 Remove your hands from capacitor to allow it to be tested 5 Read the measurement on the display CAUTION CAUTION To avoid possible damage to the meter or the equipment under test disconnect circuit power and discharge the capacitor before measuring capac...

Страница 5: ...strument 2 SET Set high low limits for compare mode 3 REC Static recording mode 4 Compare mode 5 TOL Tolerance mode 6 REL Relative mode 7 HOLD SAVE Data hold To store the setting value into the memory 8 RANGE AUTO Manual range Auto range 9 HI LO High Low limits Backlight display 8 5 4 1 2 3 9 7 6 ...

Страница 6: ... Up or Down to change the parameter 3 Press to select the digit to be adjusted the selected digit will flash 4 Press and hold for more than 1 s to save your setting 5 Press for more than 1 s to exit setup mode Features and Functions Actions Steps To power ON or OFF Press To enable data hold function Press To trigger holding next reading Press momentarily To exit data hold mode Press for more than ...

Страница 7: ...not be set when an overload value exists REL annuciator will be displayed Press To renew the relative value Press again To exit relative mode Press for more than 1 s To select manual range and to turn off the AUTO annunciator Press To step up a range at a time Press again To select auto range In auto range mode the AUTO annunciator is displayed and the instrument will select an appropriate range f...

Страница 8: ...has been used for compare mode The two right digits indicate current compare set The range from 01 to 25 The primary display shows the present measurement In this state it is ready for testing If the reading is beyond the high limit will be indicated will be indicated if the reading is out of the low limit The beeper will beep three times and the secondary display will indicate nGo If the reading ...

Страница 9: ...ting mode a To select which digit to be adjusted b To increase or decrease the current digit s value c To select High or Low limit to be set d To store the setting value in the memory The beeper will beep twice if the selected value has been stored If the current setting do not meet the rule that the high limit must be equal or greater than the low limit the beeper will beep three times e To selec...

Страница 10: ...d AS NZS CISPR11 2004 Safety Symbols CAUTION CAUTION A CAUTION notice denotes a hazard It calls attention to an operating procedure practice or the like that if not correctly performed or adhered to could result in damage to the product or loss of important data Do not proceed beyond a CAU TION notice until the indi cated conditions are fully understood and met WARNING A WARNING notice denotes a h...

Страница 11: ...This information is subject to change without notice Keysight Technologies 2009 2012 2014 Edition 3 August 2014 U1701 90047 U1701 90047 www keysight com ...

Отзывы: