General Specifications

Specifications

Temperature Adapter

Operating Temperature

0 °C to 55 °C (14 °F to 131 °F)

Storage Temperature

–20 °C to 70 °C (–4 °F to 158 °F)

Relative Humidity

Max 80% RH for temperature up to 35 °C decreasing linearly to 50% RH at 55 °C

Battery life

1600 hours approximately based on alkaline battery

Altitude

2000 meters

EMC Compliance

Susceptibility and Emissions: Commercial Limits per EN61326-1

Dimensions (HxWxD)

31 mm (H) x 59 mm (W) x 113.5 mm (L)

Weight

120 grams

Power Supply:

Single standard 9 V Battery can use Alkaline or Carbon-zinc.

Battery Types

ANSI/NEDA

IEC

Alkaline

1604A

6LR61

Carbon-zinc

1604D

6F22

Electrical Specifications 

Range

Resolution

Accuracy        

± (% of r no. of least significant digits) at 23 °C  ± 5 °C, with 

relative humidity less than 80% R.H.

–50 °C ~ 1000 ° C

1 mV/°C

–50 ~ –21 °C

2.5% + 2 °C

–20 ~ 350 °C

0.5% + 2 °C

351 ~ 500 °C

1.75% + 2 °C

501 ~ 1000 °C

2% + 2 °C

–58 °F ~ 1832 °F

1 mV/°F

–58 ~ –5.8 °F

2.5% + 3.6 °F

–4 ~ 662 °F

0.5% + 3.6 °F

664 ~ 932 °F

1.75% + 3.5 °F

933 ~ 1832 °F

2% + 3.6 °F

• The accuracy specifications does not include the error of the probe or the digital multimeter, and this 

   adapter should be stabilized for at least one hour on the DUT to be measured. Please refer to the probe 

   accuracy specifications for additional details.

• Accuracy enhancement for 351 °C ~ 500 °C, subtract 3 degrees from the reading. The accuracy is now 

   0.75% ±2 °C.

• Accuracy enhancement for 663 °F  ~ 932 °F, subtract 5.4 degrees from the reading. The accuracy is now 

   0.75% ±3.6 °F.

Operation

Follow the procedures provided to select Celsius or Fahrenheit temperature unit in proportion to mili-volt 

output using slider switch:

1. 

Ensure the temperature adapter is stabilized in operating environment.

2. 

Switch the adapter slider’s switch to set °C or °F.

3. 

Plug the banana plugs into V/COM terminals on multimeter.

4. 

Set the DC voltage measurement and adequate range on multimeter or oscilloscope in use.

5. 

Plug the miniature K-Type thermocouple probe into adapter.

6. 

Observe the DC mili-volt on a multimeter which is proportional to the temperature. (1 mV is 1 °C or ° F)

7.       Select a better resolution to 0.1 mV on a multimeter if want to see the difference to 0.1 degree.

Calibration Equipment

         

The pre-calibration guide lines are shown as follows:

 

Be sure you  are a qualified person to perform the calibration

 

The environment should be 23 °C ±2 °C and the relative humidity (RH) shall be < 80%.

The test equipment requirements listed in table below or equivalents are required to perform the calibration 

and performance verification test procedures. Alternative equipment may be used as long as the accuracy is as 

good as or better than the specifications listed.