Keysight U1586B Temperature Adapter 

Operating Instructions

The U1586B temperature adapter is designed for Keysight handheld digital multimeter (DMM), Keysight 

handheld oscilloscope, and K-Type thermocouple.

Assistance

For technical assistance, contact your nearest Keysight Sales Office or visit the Keysight website at 

www.keysight.com/find/assist 

for further information.

Safety Informations

Please use the Keysight U1586B temperature adapter only as specified in this manual. Otherwise, the protec

-

tion provided may be impaired. Refer to the safety information below.

 

 

 

 

          

 

WARNING

To avoid possible electric shock, personal injury or damage to this instrument and ensure that you use the 

adapter safely, follow these guidelines:

- Do not use the adapter if it is damaged. Before you use the adapter inspect the case. Look for cracks or 

missing 

  plastic. Pay particular attention to the insulation surrounding the connectors.

- Do not touch thermocouple probe to hazardous voltage.

- Do not operate the adapter around explosive gas, vapor or dust.

- When servicing the adapter, use only specified replacement parts.

- Removed all connections as battery replacement

- Use caution when working above 30 V ac rms, 42 V peak or 60 V dc. Such voltage pose a shock hazard.

- Avoid working alone.

- Do not operate the adapter as the cover removed or loosened.

     

CAUTION

To avoid possible damage to the temperature adapter or to the equipment under test, follow below guide

-

lines:

- Do not  connect output banana plugs to any power source.

- Use the proper terminals, function, and range for your measurements.

AC - Alternating Current

Caution, risk of danger (Refer to the user’s 

and service guide for details)

Double Insulation

DC - Direct Current

Ground

Battery Voltage Test

Set the adapter’s slider switch to OFF position, the output will output the proportional mili-volt to represent 

the battery voltage inside. The proportional scale is DC 9.9 mV per 1 V of battery voltage inside. The operat

-

ing voltage suggested is up to 7.5 V of battery. So, always make sure the mili-volt output is up to 74.2 mV 

before temperature measurement.

Battery Voltage

10 V

9 V

8 V

7.5 V

Output of U1586B

99.0 mV

89.1 mV

79.2 mV

74.2 mV

Introductions

The U1586B temperature adapter accepts any K-Type thermocouple with miniature connector  and converts 

it to 1 milli-volt per degree (Celsius or Fahrenheit).

Standard Item Purchased Checklist   

      

The following items are included when you make a purchase:

 

U1586B temperature adapter

 

K-Type bead probe

 

Operation Instructions sheet (this sheet)

 

Conform to European Union Directives 

           IEC/ EN 61010-12001 (2nd Edition)

 

OUTPUT: mV output proportional 

to inside battery. °C and °F as the 

position of slider switch

Slider switch: 

• OFF: inside battery level

• °C: 1 mV/°C

• °F: 1 mV/°F

INPUT: K-type thermocouple 

with miniature connector